Increased Peak Inspiratory Pressure Due to Intraluminal Bulging of the Inner Layer of the Reinforced Wire Tube during Anesthesia: A case report

Affiliations
  • 1Department of Anesthesiology, Yonsei University College of Medicine, Seoul, Korea.

Abstract

Excessive peak airway pressure during general endotracheal anesthesia may result from bronchospasm due to light anesthesia or surgical stimulation, bronchial intubation, tension pneumothorax, pulmonary edema, or mechanical obstruction of tube, whether from kinking, inspissated secretions, or overinflation of the cuff. Usually these problems are differentiated with auscultation and drug administration. However, mechanical problems associated with the endotracheal tube may be a cause of increased airway pressure. Reinforced, anode, or armored tubes consist of two coatings of latex or PVC that enclose spiral metal windings. Because of that, the inner layer may peel away, and intraluminally bulge due to nitrous oxide and cause airway obstruction during the course of an anesthetic process. We report a case of intraluminal bulging of the inner layer in a reinforced tube using fiberoptic bronchoscopy during anesthesia.

Keyword

obstruction; peak airway pressure; anode; armored reinforced wire tube

MeSH Terms

Airway Obstruction
Anesthesia*
Auscultation
Bronchial Spasm
Bronchoscopy
Electrodes
Intubation
Latex
Nitrous Oxide
Pneumothorax
Pulmonary Edema
Wind
Latex
Nitrous Oxide
Full Text Links
  • KJAE
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr