J Korean Breast Cancer Soc.  2003 Mar;6(1):24-28. 10.4048/jkbcs.2003.6.1.24.

Decrease in Pulse Oximeter Readings Following Injection of Isosulfan Blue Dye

Affiliations
  • 1Department of Surgery, College of Medicine, Yeungnam University, Deagu, Korea. crystallee@med.yu.ac.kr
  • 2Department of Anesthesiology, College of Medicine, Yeungnam University, Deagu, Korea.

Abstract

PURPOSE
A sentinel lymph node mapping with blue dye has been well accepted as a common procedure in breast cancer surgery. However, it is well known that blue dye absorbed into the circulation may interfere pulse oximetery reading. The aim of this study was to evaluate the change of pulse oximetery reading by isosulfan blue dye injection during sentinel lymph node mapping. METHODS: Thirteen breast cancer patients with normal preoperative cardiopulmonary functions were studied. Four ml of isosulfan blue dye was injected subdermally when the patient became stable after induction of general anesthesia. The pulse oximetery was monitored continuously. Multiple arterial blood gas analyses (ABGA) were performed before dye injection and 10, 30, 40 minutes after dye injection. The results of oxygen saturturation by oximetery (SpO2) and the results of arterial oxygen tension (SaO2) and arterial oxygen saturation (SaO2) by ABGA were compared. RESULTS: The value of both SaO2 and PaO2 measured by ABGA has not been altered by isosulfan dye injection. However SpO2 decreased by isosulfan dye injection. SpO2 decrease started 8.2+/-1.5 (2~0) minutes after dye injection and returned to preinjection level by 85.7+/-5.6 (60~126) minutes after injection. The lowest vaule of SpO2 was 95.6+/-1.2% (93~97). Mean duration of SpO2 decrease was 77.5+/-6.2 (40~117) minutes. The duration of SpO2 decrease was longer in the aged patients, but it was not statistically significant (p=0.3). There was no siginificant difference in duration of SpO2 decrease according to injection site, operation method, and body mass index (BMI). CONCLUSION: .Isosulfan dye injection using for sentinel lymph node mapping causes no change in true ABGA results but causes a mild reversible decrease in SpO2, It is important to look for other causes when SpO2 decrease is significant and persistent.

Keyword

Breast cancer; Sentinel node; Isosulfan blue dye; Pulse oximeter

MeSH Terms

Anesthesia, General
Blood Gas Analysis
Body Mass Index
Breast Neoplasms
Humans
Lymph Nodes
Oxygen
Reading*
Oxygen
Full Text Links
  • JKBCS
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr