J Lab Med Qual Assur.
2004 Jun;26(1):171-176.
Diagnosis of Malaria Using Automatic Hematology Analyzer
- Affiliations
-
- 1Department of Laboratory Medicine, Wonju Medical Center, Korea. sjjajang@paran.com
- 2Department of Laboratory Medicine, Kangwon National University College of Medicine, Korea.
- 3Department of Laboratory Medicine, Korea University College of Medicine, Korea.
Abstract
-
BACKGROUND: For the diagnosis of malaria, examination of blood smear slides by light microscopy is used as standard, and commercial kits detecting malarial antibodies and antigens are available, and molecular methods such as polymerase chain reaction (PCR) are used additionally. But, these diagnostic methods can be performed when clinicians request them, so problems of misdiagnosing the patients who are not suspected malaria may be occurred.
METHODS
In 42 Korean patients with malaria, the author analyzed the characteristic signals of malaria using granularity (90 degrees depolarized) versus lobularity (90 degrees polarized) graph of Cell-Dyn 4000 (CD4000) automatic hematologic analyzer. And, the author examined the presence of malaria in 421 random samples by CD4000 and Giemsa stain.
RESULTS
The usefulness of CD4000 in diagnosing malaria are as follows, 93.0% sensitivity, 99.3% specificity, 93.0% positive-predictive value, and 99.3% negative-predictive value.
CONCLUSION
CD4000 automatic hematology analyzer has high diagnostic sensitivity and specificity in diagnosing malaria. Because complete blood count (CBC) is the routine test for most patients, this method has advantage of time and cost effectiveness and can even detect malaria in unsuspected cases.