Asian Spine J.  2011 Mar;5(1):51-58. 10.4184/asj.2011.5.1.51.

Reliability of Triggered EMG for Prediction of Safety during Pedicle Screw Placement in Adolescent Idiopathic Scoliosis Surgery

Affiliations
  • 1Department of Orthopaedic Surgery, Kyungpook National University Hospital, Kyungpook National University School of Medicine, Daegu, Korea. wkmin@knu.ac.kr
  • 2Department of Physiology, Kyungpook National University School of Medicine, Daegu, Korea.

Abstract

STUDY DESIGN: We performed a prospective study to evaluate the reliability of using triggered electromyography (EMG) for predicting pedicle wall breakthrough during the placement of pedicle screw in adolescent idiopathic scoliosis surgery. PURPOSE: We wanted to correlate pedicle wall breakthrough with the triggered EMG threshold of stimulation and the postoperative computed tomography (CT) findings. OVERVIEW OF LITERATURE: Pedicle wall breakthrough has been reported to be difficult to evaluate by radiographs. Triggered EMG had been found to be a more sensitive test to detect this breakthrough.
METHODS
Seven patients who underwent the insertion of 103 pedicle screws were evaluated. The triggered EMG activity was recorded from several muscles depending on the level of screw placement. The postoperative CT scans were read by a spine surgeon who was a senior fellow in orthopedics, and a musculoskeletal radiologist.
RESULTS
The mean age at surgery was 12.6 years (range, 11 to 17 years). The preoperative mean Cobb angle was 54.7degrees (range, 45 to 65degrees). There were 80 thoracic screws and 23 lumbar screws. All the screws had stimulation thresholds of > or = 6 mA, except 3 screws with the stimulation threshold of < 6 mA. Ten screws (9.7%) showed violation of the pedicle wall on the postoperative CT scans. Five screws penetrated medially and another five penetrated laterally. No postoperative neurologic complications were noted in any of the seven patients.
CONCLUSIONS
Measuring the stimulation threshold of triggered EMG helps to assess the pedicle screw placement. Pedicle screws that had stimulation threshold of > or = 6 mA were safe, with 90.3% reliability, as was assessed on the postoperative CT scans.

Keyword

Adolescent idiopathic scoliosis; Pedicle screw; Triggered electromyography

MeSH Terms

Adolescent
Electromyography
Humans
Muscles
Orthopedics
Prospective Studies
Scoliosis
Spine
Full Text Links
  • ASJ
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr