J Korean Soc Magn Reson Med.
1997 Dec;1(1):66-74.
Quantitative Analysis of Susceptibility Effects in TRFGE and CGE Sequences for Functional MRI
- Affiliations
-
- 1Department of Electrical Engineering, KAIST, Seoul.
- 2Department of Computer Engineering, Teajon University, Teajon.
Abstract
- fMRI, functional MRI introduced receently appears based on the gradient echo technique which is sensitive to the field inhomogeneity developed due to the local susceptibility changes of blood oxygenation and deoxygenation. There has been many variants of the basic gradient echo sequence which is sensitive to the local inhomogeniety, among others such as GRASS or SSFP to EPISTAR are the most commonly used gradient echo techniques. Common to all these gradient echo techniques is that the signal due to the susceptibility effects is generally decreased with increasing inhomogeneity due to the T2 effect or conventionally konwn as blood oxygenation level dependent(BOLD) effect. It is, also found that the BOLD sensitivity is also dependent on the imaging modes, namely whether the imaging is in axial, or coronal or sagittal mode as well as the directions of the vessels against the main magnetic field. We have, therefore, launched a systematic study of imaging mode dependent signal change or BOLD sensitivity as well as the signal changes due tothe tilting angle of the imaging planes. Study has been made for both TRFGE sequence and CGE sequence to compare the distinctions of the each mode since each technique has different sensitivity againsst susceptibility effect. Method of computation and both the computer simulations and their corresponding experimental results are presented.