Korean J Anesthesiol.  2013 May;64(5):407-413. 10.4097/kjae.2013.64.5.407.

Impact of priming the infusion system on the performance of target-controlled infusion of remifentanil

Affiliations
  • 1Department of Anesthesiology and Pain Medicine, Ajou University School of Medicine, Suwon, Korea. anesmin@nate.com
  • 2Department of Anesthesiology and Pain Medicine, Gachon University Gil Medical Center, Incheon, Korea.

Abstract

BACKGROUND
The start-up behavior of syringe and syringe pump is known to be one of the causes of inaccurate intravenous infusion. This study evaluated the method of priming the infusion system (PRIMING), and its impact on the target-controlled infusion (TCI) of two remifentanil diluents.
METHODS
PRIMING was performed using an evacuation of 2.0 ml to the atmosphere prior to TCI. Forty-eight TCI, using 50 microg/ml (Remi50) or 20 microg/ml (Remi20) of diluents, were performed targeting 4.0 ng/ml of effect-site concentration (Ceff), with PRIMING or not. The gravimetrical measurements of the delivered infusates reproduced actual Ceff. The bolus amount and time to reach 95% target were compared.
RESULTS
Without PRIMING, Remi50 infused less bolus (43 +/- 23 %) than Remi20 (19 +/- 9 %) (P = 0.003), and showed more delayed increase of Ceff (11.2 +/- 4.0 min) than Remi20 (7.4 +/- 0.4 min) (P = 0.028). However, PRIMING significantly decreased the deficit of the bolus (2 +/- 1%), as well as the delay of the increase of Ceff in Remi50 (1.2 +/- 0.2 min) (both P < 0.001). In addition, with PRIMING, the start-up bolus showed minimal difference to the nominal bolus (1 and 2%), and Ceff were increased to 4.0 +/- 0.1 ng/ml at the expected time of peak effect, irrespective of the diluents.
CONCLUSIONS
Proper operation of the syringe pump used in the priming of the syringe may be helpful in reduction of the inaccuracy of TCI, particularly during the early phase of infusion, or the infusion of a more concentrated diluent.

Keyword

Drug delivery; Priming; Target-controlled infusion

MeSH Terms

Atmosphere
Infusions, Intravenous
Piperidines
Syringes
Piperidines

Cited by  1 articles

Correction of target-controlled infusion following wrong selection of emulsion concentrations of propofol
Yun-Jeong Chae, Han Bum Joe, Won-Il Lee, Jin-A Kim, Sang-Kee Min
Korean J Anesthesiol. 2014;66(5):377-382.    doi: 10.4097/kjae.2014.66.5.377.

Full Text Links
  • KJAE
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr