J Korean Med Sci.  1994 Dec;9(6):505-506. 10.3346/jkms.1994.9.6.505.

Bilateral fractures of Hydroflex penile prostheses: an unusual complication

Affiliations
  • 1Department of Urology, College of Medicine, Chung-Ang University, Seoul, Korea.

Abstract

We report bilaterally fractured Hydroflex implants in 2 patients. One fracture was complete and three were incomplete. All the fractures occurred at the junction of the rear reservoir and the inflation chamber.

Keyword

Hydroflex penile prostheses; fracture

MeSH Terms

Adult
Case Report
Equipment Design
Human
Male
*Penile Prosthesis
Prosthesis Failure