Imaging Sci Dent.  2025 Mar;55(1):37-47. 10.5624/isd.20240182.

Effect of a prototype 2-dimensional antiscatter grid on image quality obtained with a dental cone-beam computed tomography scanner

Affiliations
  • 1M2 Technologies, Denver, CO, USA
  • 2Department of Radiology, Stanford University, Stanford, CA, USA
  • 3Planmeca, Helsinki, Finland
  • 4Department of Physics, School of Natural Sciences, Technical University of Munich, Garching, Germany
  • 5Munich Institute of Biomedical Engineering, Technical University of Munich, Garching, Germany
  • 6Department of Radiation Oncology, University of Colorado School of Medicine, Aurora, CO, USA
  • 7Department of Diagnostic and Biomedical Sciences, UTHealth Houston School of Dentistry, Houston, TX, USA

Abstract

Purpose
X-ray scattering adversely affects cone-beam computed tomography (CBCT) image quality, generating image artifacts, causing inaccurate tissue density representation, and reducing contrast. This study evaluated the performance of a 2-dimensional antiscatter grid (2D grid) prototype in a dental CBCT system.
Materials and Methods
A focused 2D grid prototype was fabricated from tungsten and integrated with the detector of a dental CBCT system. Residual scatter transmitted through the 2D grid was corrected using a measurement-based scatter correction method. Phantom imaging experiments were performed in anatomical regions relevant to dental and head imaging with and without this grid. Following image reconstruction via filtered back projection, attenuation coefficients were converted to Hounsfield units (HU). Subsequently, scatter suppression performance, HU consistency, image artifacts, and contrast resolution were evaluated.
Results
The 2D grid reduced scatter intensity by a factor of 10-20 in CBCT projections. Consequently, the grid substantially increased contrast, reduced image artifacts, and improved HU consistency. The contrast increased by 27% and 48% in bone- and soft tissue-equivalent regions, respectively. HU value deviations among teeth decreased from 510 to 146 HU. These results indicate improved visualization and tissue density representation fidelity in CBCT images acquired with the 2D grid.
Conclusion
Use of a 2D grid could substantially improve the accuracy of tissue density representation and the contrast of dental and head anatomy in 3-dimensional images obtained with dental CBCT. Such improvements may translate to better quantitative evaluation of bone quality, enhanced tissue visualization, and more accurate model generation for surgical planning and guidance.

Keyword

Cone-Beam Computed Tomography; X-rays; Diagnostic Imaging; Image quality enhancement