Clin Orthop Surg.  2025 Feb;17(1):166-173. 10.4055/cios24190.

Reliability of Ultrasound Elastography According to Experience Level and Anatomic Location

Affiliations
  • 1Department of Orthopedic Surgery, Korea University Anam Hospital, Korea University College of Medicine, Seoul, Korea

Abstract

Background
Strain elastography (SE) and shear wave elastography (SWE) are emerging techniques for evaluating the elasticity of soft tissue. This study aimed to determine interobserver and intraobserver reliability for elasticity measurements of different tissues and anatomic locations using SE and SWE.
Methods
Ten healthy adult male individuals with 20 upper extremities participated in this study. The elasticities of the wrist extensor muscle, the common extensor tendon, and supraspinatus tendon were measured. Strain ratio and shear wave velocity were measured twice by 2 different examiners (examiner 1 with over 20 years of experience in musculoskeletal sonography and examiner 2 with 1 year of experience). Interobserver and intraobserver reliability was assessed using the intraclass correlation coefficient (ICC).
Results
The 10 individuals’ age ranged from 28 to 35 years. In SE, interobserver reliabilities at the 3 anatomic locations (wrist extensor muscle, common extensor tendon, and supraspinatus tendon) showed fair to moderate agreement (ICC = 0.489, p = 0.076; ICC = 0.408, p = 0.131; and ICC = 0.296, p = 0.711, respectively). The intraobserver reliabilities of examiner 1 were moderate to substantial only at the wrist extensor muscle and the common extensor tendon (ICC = 0.563, p = 0.039 and ICC = 0.702, p = 0.006, respectively). In SWE, interobserver reliabilities for the wrist extensor muscle and the supraspinatus tendon were moderate to substantial (ICC = 0.756, p = 0.002 and ICC = 0.565, p = 0.039, respectively). The intraobserver reliabilities of examiner 1 at the 3 anatomic locations were almost perfect (ICC = 0.843, p = 0.001; ICC = 0.800, p = 0.001; and ICC = 0.825, p = 0.001, respectively). The results of examiner 2 showed almost perfect agreement at the wrist extensor muscle (ICC = 0.886, p = 0.001) and moderate to substantial agreement at the tendons of the common extensor and supraspinatus (ICC = 0.592, p = 0.029 and ICC = 0.682, p = 0.008, respectively).
Conclusions
SWE is a reliable method for assessing the flexibility of soft tissue, but it is affected by expertise and the specific anatomical site.

Keyword

Elasticity imaging techniques; Ultrasonography; Elbow; Shoulder; Rotator cuff