Unexpected Restart Failure of Durable Left Ventricular Assist Devices: A Report of Two Cases

Affiliations
  • 1Department of Thoracic and Cardiovascular Surgery Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Korea
  • 2Division of Cardiology, Department of Medicine, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Korea

Abstract

The HeartWare Ventricular Assist Device (HVAD) was widely used for mechanical circulatory support in patients with end-stage heart failure. However, there have been reports of a critical issue with HVAD pumps failing to restart, or experiencing delays in restarting, after being stopped. This case report describes 2 instances of HVAD failure-to-restart during heart transplantation surgery and routine outpatient care. Despite multiple attempts to restart the pump using various controllers and extensions, the HVAD failed to restart, triggering a hazard alarm for pump stoppage. In one case, the patient survived after receiving a heart transplantation, while in the other, the patient died immediately following the controller exchange. These cases highlight the rare but life-threatening complication of HVAD failure-to-restart, underscoring the importance of awareness among clinicians, patients, and caregivers, and adherence to the manufacturer’s guidelines and recommendations for HVAD management.

Keyword

Heart failure; Cardiomyopathies; Heart-assist devices; Heart transplantation; Extracorporeal membrane oxygenation; Case reports
Full Text Links
  • JCS
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr