Background
The purpose of this study was to evaluate the clinical and radiological outcomes of high-flexion total knee arthroplasty (TKA) using Vega Knee System (B. Braun, Aesculap) at a long-term follow-up and to analyze the implant survivorship.
Methods
We enrolled 165 patients (232 knees) with a minimum 7-year follow-up after TKA (VEGA Knee System). For clinical assessment, range of motion (ROM), Knee Injury and Osteoarthritis Outcome Score (KOOS), and Western Ontario and McMaster University Osteoarthritis Index (WOMAC) were used. For radiologic assessment, hip-knee-ankle angle, component position, and the existence of radiolucent lines and loosening were used. Survival analysis was conducted using the Kaplan-Meier method.
Results
The mean follow-up period was 9.8 years. The mean ROM increased from 124.4° to 131.4° at the final follow-up. The WOMAC score decreased from 38.5 to 17.4 at the final follow-up (p < 0.001). All 5 subscales of the KOOS improved at the final follow-up (all subscales, p < 0.001). Revision TKA was performed in 10 cases (4.3%), which included 9 cases of aseptic loosing and 1 case of periprostatic joint infection. Of the 9 aseptic loosening cases (3.9%), 8 cases (3.4%) were loosening of the femoral component and 1 case (0.4%) was loosening of the tibial component. When revision for any reason was considered an endpoint, the 10-year survivorship was 96.2% (95% confidence interval [CI], 93.9%–98.5%). On the other hand, when revision for aseptic loosening was considered the endpoint, the 10-year survivorship was 96.6% (95% CI, 94.4%–98.8%).
Conclusions
The Vega Knee System provided good clinical results in the long-term follow-up period. Although the VEGA Knee System showed acceptable implant survivorship, loosening of the femoral component occurred in about 3.4% of the patients. For more accurate evaluation of the survivorship of high-flexion design TKA with a short posterior flange, it is necessary to conduct more long-term follow-up studies targeting diverse races, especially Asians who frequently perform high-flexion activities.

Keyword

Total knee arthroplasty; High-flexion; Treatment outcome; Survivorship; Prosthesis failure