J Gastric Cancer.  2023 Jul;23(3):451-461. 10.5230/jgc.2023.23.e22.

Efficacy and Safety of ClearCut™ Knife H-type in Endoscopic Submucosal Dissection for Gastric Neoplasms: A Multicenter, Randomized Trial

Affiliations
  • 1Department of Internal Medicine, Hallym University College of Medicine, Chuncheon, Korea
  • 2Division of Gastroenterology, Department of Internal Medicine, Hallym University Sacred Heart Hospital, Hallym University College of Medicine, Anyang, Korea
  • 3Department of Internal Medicine, Gangneung Asan Hospital, University of Ulsan College of Medicine, Gangneung, Korea
  • 4Department of Gastroenterology, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Korea

Abstract

Purpose
Endoscopic submucosal dissection (ESD) is an effective treatment for early gastrointestinal neoplasms. However, this is a time-consuming procedure requiring various devices. This study aimed to evaluate the efficacy and safety of the ClearCut™ Knife H-type, which is an integrated needle-tipped and insulated-tipped (IT) knife.
Materials and Methods
Between July 2020 and September 2021, 99 patients with gastric epithelial neoplasms scheduled for ESD at three tertiary care hospitals were randomly assigned to H-knife (ClearCut™ Knife H-type) or IT-knife (conventional IT knife) groups. Procedure times, therapeutic outcomes, and adverse events were analyzed.
Results
A total of 98 patients (50 in the H-knife group and 48 in the IT-knife group) were analyzed. The median total procedure time was 11.9 minutes (range, 4.4–47.2 minutes) in the H-knife group and 12.7 minutes (range, 5.2–137.7 minutes) in the IT-knife group (P=0.209). Unlike the IT-knife group, which required additional devices in all cases, no additional devices were used in the H-knife group (P<0.001). En-bloc resection was performed for all lesions in both groups. The incidence of adverse events was not significantly different between groups (4.0% in the H-knife group vs. 8.3% in the IT-knife group; P=0.431).
Conclusions
The newly developed hybrid device, the ClearCut™ Knife H-type, had comparable efficacy to the conventional IT knife for gastric ESD.

Keyword

Endoscopic submucosal dissection; Medical devices; Treatment outcome
Full Text Links
  • JGC
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles