Nucl Med Mol Imaging.  2023 Jun;57(3):126-136. 10.1007/s13139-022-00787-x.

Dual‑Tracer Parathyroid Imaging Using Joint SPECT Reconstruction

Affiliations
  • 1Biomedical Engineering Program, Faculty of Engineering, Chulalongkorn University, Bangkok, Thailand
  • 2Chulalongkorn University Biomedical Imaging Group, Faculty of Medicine, Department of Radiology, Chulalongkorn University, Bangkok, Thailand
  • 3Division of Nuclear Medicine, Faculty of Medicine, Department of Radiology, King Chulalongkorn Memorial Hospital, Thai Red Cross Society, Chulalongkorn University, Bangkok, Thailand
  • 4Division of Nuclear Medicine, Faculty of Medicine, Department of Radiology, Chulalongkorn University, Bangkok, Thailand

Abstract

Purpose
We assessed the lesion detection performance of the dual-tracer parathyroid SPECT imaging using the joint reconstruction method.
Materials and Methods
Thirty-six noise realizations were created from SPECT projections collected from an in-house neck phantom to emulate 99mTc-pertechnetate/99mTc-sestamibi parathyroid SPECT datasets. Difference images representing parathyroid lesions were reconstructed using the subtraction and the joint methods whose corresponding optimal iteration was defined as the iteration which maximized the channelized Hotelling observer signal-to-noise ratio (CHO-SNR). The joint method whose initial estimate was derived from the subtraction method at optimal iteration (the joint-AltInt method) was also assessed. In a study of 36 patients, a human-observer lesion-detection study was performed using difference images from the three methods at optimal iteration and the subtraction method with four iterations. The area under the receiver operating characteristic curve (AUC) was calculated for each method.
Results
In the phantom study, both the joint-AltInt method and the joint method improved SNR compared to the subtraction method at their optimal iteration by 444% and 81%, respectively. In the patient study, the joint-AltInt method yielded the highest AUC of 0.73 as compared with 0.72, 0.71, and 0.64 from the joint method, the subtraction method at optimal iteration, and the subtraction method at four iterations. At a specificity of at least 0.70, the joint-AltInt method yielded significantly higher sensitivity than the other methods (0.60 vs 0.46, 042, and 0.42; p < 0.05).
Conclusions
The joint reconstruction method yielded higher lesion detectability than the conventional method and holds promise for dual-tracer parathyroid SPECT imaging.

Keyword

Parathyroid imaging; Dual-tracer technique; Image reconstruction; SPECT