Imaging Sci Dent.  2022 Sep;52(3):245-258. 10.5624/isd.20220019.

CBCT-based assessment of root canal treatment using micro-CT reference images

Affiliations
  • 1Department of Restorative Dentistry, School of Dentistry of Ribeirão Preto, University of São Paulo, Brazil
  • 2OMFS IMPATH Research Group, Department of Imaging and Pathology, Faculty of Medicine, University of Leuven, University Hospitals Leuven, Leuven, Belgium
  • 3Faculty of Dentistry, University of Ribeirão Preto, Ribeirão Preto, Brazil
  • 4Department of Radiology, Faculty of Dentistry, Chulalongkorn University, Bangkok, Thailand
  • 5Aarhus Institute of Advanced Studies, Aarhus University, Aarhus, Denmark
  • 6Department of Radiology, Universitair Ziekenhuis Brussel, Vrije Universiteit Brussel, Brussels, Belgium
  • 7Department of Dental Medicine, Karolinska Institutet, Stockholm, Sweden

Abstract

Purpose
This study compared the root canal anatomy between cone-beam computed tomography (CBCT) and micro-computed tomography (micro-CT) images before and after biomechanical preparation and root canal filling.
Materials and Methods
Isthmus-containing mesial roots of mandibular molars (n = 14) were scanned by micro-CT and 3 CBCT devices: 3D Accuitomo 170 (ACC), NewTom 5G (N5G) and NewTom VGi evo (NEVO). Two calibrated observers evaluated the images for 2-dimensional quantitative parameters, the presence of debris or root perforation, and filling quality in the root canal and isthmus. The kappa coefficient, analysis of variance, and the Tukey test were used for statistical analyses (α= 5%).
Results
Substantial intra-observer agreement (κ= 0.63) was found between micro-CT and ACC, N5G, and NEVO. Debris detection was difficult using ACC (42.9%), N5G (40.0%), and NEVO (40%), with no agreement between micro-CT and ACC, N5G, and NEVO (0.05<κ<0.12). After biomechanical preparation, 2.4%-4.8% of CBCT images showed root perforation that was absent on micro-CT. The 2D parameters showed satisfactory reproducibility between micro-CT and ACC, N5G, and NEVO (intraclass correlation coefficient: 0.60-0.73). Partially filled isthmuses were observed in 2.9% of the ACC images, 8.8% of the N5G and NEVO images, and 26.5% of the micro-CT images, with no agreement between micro-CT and ACC, and poor agreement between micro-CT and N5G and NEVO. Excellent agreement was found for area, perimeter, and the major and minor diameters, while the roundness measures were satisfactory.
Conclusion
CBCT images aided in isthmus detection and classification, but did not allow their classification after biomechanical preparation and root canal filling.

Keyword

Cone-Beam Computed Tomography; X-Ray Microtomography; Endodontics