Abstract

Background
Lospa posterior-stabilized (PS) Plus type is a modified version of Lospa PS, in which the polyethylene insert shape is modified to reinforce stability and prevent patella-post impingement compared to Lospa PS. However, studies comparing the clinical and radiographic results of the two designs have not been reported yet. This study aimed to compare the clinical results of total knee arthroplasty (TKA) using the existing PS type and the modified Lospa PS Plus type.
Methods
A retrospective study was performed on 558 knees of 342 patients who underwent TKA using the Lospa PS or PS Plus types and were followed up for at least 2 years. Cases were divided into two groups according to the implant used: 212 cases in the PS group and 346 cases in the PS Plus group. For clinical outcome assessment, knee range of motion (ROM), Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC) score, and Knee Society Score (KSS) were recorded before surgery and at the 2-year follow-up. Radiographic outcomes were evaluated according to the American Knee Society method. The incidence of postoperative complications and survival rates were compared between the two groups.
Results
Both groups showed significant clinical improvement after surgery. The average KSS significantly improved from 53.4 points in the PS group and 52.3 points in the PS Plus group preoperatively to 91.3 points and 93.2 points after surgery, respectively (p < 0.001). The average WOMAC score improved from 50.4 points in the PS group and 52.3 points in the PS Plus group before surgery to 15.6 points and 14.8 points after surgery, respectively (p < 0.001). There was no significant difference between the two groups in ROM, the alignment of the lower limbs, and the implant position after surgery. The complication rates were also similar between the groups (p = 0.167).
Conclusions
The Lospa PS Plus model is a modified design that improves the post structure from the previous PS type. Compared to the PS type, the PS Plus type showed similar statistical results at 2-year follow-up and good clinical results. The short-term average survival rate was over 98%, showing promising results.

Keyword

Knee; Total knee arthroplasty; Posterior-stabilized