Ultrasonography.  2022 Jan;41(1):164-170. 10.14366/usg.21071.

Controlled attenuation parameter measured using transient elastography for the noninvasive assessment of macrovesicular steatosis in potential living liver donors

Affiliations
  • 1Department of Radiology and Research Institute of Radiological Science, Severance Hospital, Yonsei University College of Medicine, Seoul, Korea
  • 2Department of Radiology and Research Institute of Radiology, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Korea
  • 3Department of Surgery, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Korea

Abstract

Purpose
This study aimed to determine the diagnostic performance of the controlled attenuation parameter (CAP) measured using transient elastography (TE) for assessing macrovesicular steatosis (MaS) in potential living liver donors using same-day biopsy as a reference standard.
Methods
This retrospective study included 204 living liver donor candidates who underwent TE and liver biopsy on the same day between July 2013 and June 2014. The histologic degree of MaS was determined. The area under the receiver operating characteristic curve (AUROC) was used to evaluate the performance of CAP for diagnosing MaS of >10%, and the optimal cutoff value was identified using the maximal Youden index.
Results
Based on liver biopsy, 185 subjects had MaS of ≤10% and 19 had MaS of >10%. The CAP value was significantly correlated with the percentage of MaS on liver biopsy (r=0.635, P<0.001), and the median CAP value was significantly higher in subjects with MaS of >10% than in those with MaS of ≤10% (300 dB/m vs. 209 dB/m, P<0.001). The AUROC for diagnosing MaS of >10% by CAP was 0.938 (95% confidence interval, 0.896 to 0.967), and a CAP of >259 dB/m yielded a sensitivity of 84.2% and a specificity of 92.4%.
Conclusion
The CAP measured using TE was significantly correlated with MaS and accurately detected substantial MaS in potential living liver donors. The CAP is a promising tool for the noninvasive diagnosis of MaS and may be used to screen unsuitable living liver donor candidates.

Keyword

Controlled attenuation parameter; Transient elastography; Macrovesicular steatosis; Liver biopsy; Living donors