Asian Spine J.  2021 Dec;15(6):728-738. 10.31616/asj.2020.0400.

Multimodal Intraoperative Neurophysiological Monitoring in Spine Surgeries: The Experience at a Spine Centre through Years

Affiliations
  • 1Department of Anaesthesia, Sakra World Hospital, Bengaluru, India
  • 2Institute of Neurosciences, Sakra World Hospital, Bengaluru, India
  • 3Department of Neurosciences, Sakra World Hospital, Bengaluru, India

Abstract

Study Design: Retrospective observational study. Purpose: To share our experience of multimodal intraoperative neurophysiological monitoring (IONM) used in Sakra World Hospital, Bengaluru in various spine surgeries. Overview of Literature: The development of new onset postoperative neurological deficits can be completely avoided. In order to avoid these, IONM has become a standard of care in recent times for early detection and manipulation of the surgical procedure to prevent postoperative neurological deficits.
Methods
This retrospective study was performed on 408 patients who had undergone spine surgeries with IONM during April 2014 to March 2020 at a single center. The operative report, anesthesia record, and IONM were reviewed. All the patients were reassessed for postoperative neurological deficits in the postoperative period and followed up based on the intraoperative findings and neurological deficits for 4 weeks. Signal changes in IONM were reviewed, and the obtained results were further categorized into true positive, true negative, false positive, or false negative. If changes were observed during the IONM, the patients were managed as per the algorithm.
Results
Of the 408 patients being monitored continuously during the intraoperative period, 38 showed changes in recordings, 28 developed postoperative neurological deficits, and one developed neurological deficit without any change in the IONM. Nine patients had transient neurological deficits, and the other 20 had permanent neurological deficits. Overall, the multimodal IONM used in our study had a sensitivity of 96.6%, specificity of 97.4%, a positive predictive value of 73.7%, and a negative predictive value of 99.7%.
Conclusions
Use of decision algorithm and multimodal neuromonitoring consisting of motor evoked potentials, somatosensory evoked potentials, and electromyography complement each other in the detection of neurological injury during the course the surgery, improve intraoperative care, and prevent further damage and morbidity in patients.

Keyword

Intraoperative neurophysiological monitoring; Motor evoked potentials; Somatosensory evoked potentials; Electromyogram; Spine
Full Text Links
  • ASJ
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr