Clin Orthop Surg.  2021 Mar;13(1):67-70. 10.4055/cios19182.

Reliability of 2 Smartphone Applications for Cobb Angle Measurement in Scoliosis

Affiliations
  • 1Department of Orthopedics and Traumatology, Haydarpasa Numune Education and Research Hospital, Istanbul, Turkey

Abstract

Background
As mobile technology has evolved, smartphone applications have been used for radiographic angle measurements in daily clinical practice. This study aimed to assess the reliability of 2 smartphone applications (iPinPoint and Cobbmeter) in measuring scoliosis Cobb angles compared with picture archiving and communication system (PACS) tools.
Methods
Anteroposterior whole spinal digital radiographs of 50 patients were retrospectively analyzed. Four observers measured Cobb angles of predetermined major structural curves using the tools in the PACS software and 2 smartphone applications. The inter- and intraobserver reliabilty were measured using intraclass correlation coefficients (ICC).
Results
Very good interobserver agreement was seen with PACS, iPinPoint, and Cobbmeter measurements (ICC, 0.991, 0.980, and 0.991, respectively). Intraobserver reliability of the 4 observers was also very good for all techniques (ICC > 0.9 for all observers).
Conclusions
Both smartphone applications were reliable in measuring scoliosis Cobb angles, with reference to PACS tools. They may be useful when digital or manual mesurement tools are not available.

Keyword

Scoliosis; Smartphone; Software; Measures
Full Text Links
  • CIOS
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr