J Korean Neurosurg Soc.  2021 Jan;64(1):13-22. 10.3340/jkns.2020.0186.

Validation of Gamma Knife Perfexion Dose Profile Distribution by a Modified Variable Ellipsoid Modeling Technique

Affiliations
  • 1Department of Neurosurgery, Pusan National University Hospital, Busan, Korea
  • 2Gamma-knife Center, Haeundae Paik Hospital, Inje University, Busan, Korea
  • 3Gamma-knife Center, Pusan National University Yangsan Hospital, Yangsan, Korea
  • 4Department of Neurosurgery, Pusan National University Yangsan Hospital, Yangsan, Korea
  • 5Department of Radiology, Pusan National University Yangsan Hospital, Yangsan, Korea

Abstract


Objective
: High precision and accuracy are expected in gamma knife radiosurgery treatment. Because of the requirement of clinically applying complex radiation and dose gradients together with a rapid radiation decline, a dedicated quality assurance program is required to maintain the radiation dosimetry and geometric accuracy and to reduce all associated risk factors. This study investigates the validity of Leksell Gamma plan (LGP)10.1.1 system of 5th generation Gamma Knife Perfexion as modified variable ellipsoid modeling technique (VEMT) method.
Methods
: To verify LGP10.1.1 system, we compare the treatment plan program system of the Gamma Knife Perfexion, that is, the LGP, with the calculated value of the proposed modified VEMT program. To verify a modified VEMT method, we compare the distributions of the dose of Gamma Knife Perfexion measured by Gafchromic EBT3 and EBT-XD films. For verification, the center of an 80 mm radius solid water phantom is placed in the center of all sectors positioned at 16 mm, 4 mm and 8 mm; that is, the dose distribution is similar to the method used in the x, y, and z directions by the VEMT. The dose distribution in the axial direction is compared and analyzed based on Full-Width-of-Half-Maximum (FWHM) evaluation.
Results
: The dose profile distribution was evaluated by FWHM, and it showed an average difference of 0.104 mm for the LGP value and 0.130 mm for the EBT-XD film.
Conclusion
: The modified VEMT yielded consistent results in the two processes. The use of the modified VEMT as a verification tool can enable the system to stably test and operate the Gamma Knife Perfexion treatment planning system.

Keyword

Radiosurgery; Film dosimetry; Gamma knife; Perfection

Figure

  • Fig. 1. Collimating system of GK Perfexion with 576 positions that radioisotopes can be positioned : 16 mm (), 4 mm (), 8 mm ().

  • Fig. 2. Phantom and scanner for film measurement. A : Solid-water Phantom & film. B : Epson Scanner.

  • Fig. 3. The dose distribution obtained from the data extracted from LGP and the data calculated by the VEMT method are expressed using wolfram mathematica. Comparisons of dose profiles for the 4, 8, and 16 mm collimating configurations along the x axis (A), the y axis (B), and along the z axis (C) using VEMT (-) and LGP (). LGP : Leksell Gamma plan, VEMT : variable ellipsoid modeling technique.

  • Fig. 4. Fitting does calibration curve at the red, green channel.

  • Fig. 5. Normalized intensity analysis (C) of the treatment plan (A) and the actual film survey (B) for a virtual target in the x-y plane.


Reference

References

1. Baghani HR, Aghamiri SM, Mahdavi SR, Robatjazi M, Zadeh AR, Akbari ME, et al. Dosimetric evaluation of gafchromic EBT2 film for breast intraoperative electron radiotherapy verification. Phys Med. 31:37–42. 2015.
Article
2. Beck J, Berndt A. An algorithm for independent verification of Gamma Knife treatment plans. Med Phys. 31:2780–2784. 2004.
Article
3. Casanova Borca V, Pasquino M, Russo G, Grosso P, Cante D, Sciacero P, et al. Dosimetric characterization and use of GAFCHROMIC EBT3 film for IMRT dose verification. J Appl Clin Med Phys. 14:4111. 2013.
4. Chung JP, Oh SW, Seong YM, Chun KJ, Chung HT. An effective calibration technique for radiochromic films using a single-shot dose distribution in Gamma Knife®. Phys Med. 32:368–378. 2016.
Article
5. Elekta. Leksell Gamma Knife Perfexion - Application Introductory Training, (Dose calculation for Leksell Gamma Knife Perfexion). Available at : https://www.elekta.com/radiosurgery/leksell-gamma-knife-perfexion.
6. Hur BI, Choi BK, Sung SK, Cho WH, Cha SH, Choi CH. The variable ellipsoid modeling technique as a verification method for the treatment planning system of gamma knife radiosurgery. J Korean Neurosurg Soc. 47:128–133. 2010.
Article
7. Hur BI. Comparisons between the two dose profiles extracted from Leksell GammaPlan and calculated by variable ellipsoid modeling technique. J Korean Soc Radiol. 11:9–17. 2017.
Article
8. Jin JY, Drzymala R, Li Z. A simple method of independent treatment time verification in gamma knife radiosurgery using integral dose. Med Phys. 31:3337–3344. 2004.
Article
9. Khachonkham S, Dreindl R, Heilemann G, Lechner W, Fuchs H, Palmans H, et al. Characteristic of EBT-XD and EBT3 radiochromic film dosimetry for photon and proton beams. Phys Med Biol. 63:065007. 2018.
Article
10. Kondziolka D, Maitz AH, Niranjan A, Flickinger JC, Lunsford LD. An evaluation of the Model C gamma knife with automatic patient positioning. Neurosurgery. 50:429–431. discussion 431-432. 2002.
Article
11. Maitz AH, Wu A, Lunsford LD, Flickinger JC, Kondziolka D, Bloomer WD. Quality assurance for Gamma Knife stereotactic radiosurgery. Int J Radiat Oncol Biol Phys. 32:1465–1471. 1995.
Article
12. Marcu SM, Wu QJ, Pillai K, Weinhous MS. GammaPlan-Leksell Gamma Knife radiosurgery treatment planning verification method. Med Phys. 27:2146–2149. 2000.
Article
13. Najafi M, Geraily G, Shirazi A, Esfahani M, Teimouri J. Analysis of Gafchromic EBT3 film calibration irradiated with gamma rays from different systems: Gamma Knife and Cobalt-60 unit. Med Dosim. 42:159–168. 2017.
Article
14. Novotny J, Bhatnagar JP, Niranjan A, Quader MA, Huq Ms, Bednarz G, et al. Dosimetric comparison of the Leksell Gamma Knife Perfexion and 4C. J Neurosurg. 109 Suppl:8–14. 2008.
Article
15. Tsai JS, Engler MJ, Rivard MJ, Mahajan A, Borden JA, Zheng Z. A formalism for independent checking of Gamma Knife dose calculations. Med Phys. 28:1842–1849. 2001.
Article
16. Wu A, Lindner G, Maitz AH, Kalend AM, Lunsford LD, Flickinger JC, et al. Physics of Gamma Knife approach on convergent beams in stereotactic radiosurgery. Int J Radiat Oncol Biol Phys. 18:941–949. 1990.
Article
17. Zhang P, Dean D, Wu QJ, Sibata C. Fast verification of Gamma Knifetrade mark treatment plans. J Appl Clin Med Phys. 1:158–164. 2000.
Full Text Links
  • JKNS
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr