Korean J Intern Med.  2018 Jan;33(1):113-120. 10.3904/kjim.2016.161.

Diagnostic accuracy of manual office blood pressure measurement in ambulatory hypertensive patients in Korea

Affiliations
  • 1Department of Internal Medicine, Seoul National University Bundang Hospital, Seongnam, Korea. djchoi@snubh.org
  • 2Department of Internal Medicine, National Medical Center, Seoul, Korea.
  • 3Department of Internal Medicine, Seoul National University Hospital, Seoul, Korea.
  • 4Department of Internal Medicine, Hanyang University Seoul Hospital, Seoul, Korea.
  • 5Department of Internal Medicine, Severance Hospital, Yonsei University College of Medicine, Seoul, Korea.

Abstract

BACKGROUND/AIMS
Currently, office blood pressure (OBP) is the most widely used method of measuring blood pressure (BP) in daily clinical practice. However, data on the diagnostic accuracy of OBP in reference to ambulatory blood pressure (ABP) are scarce in Korea.
METHODS
In retrospective and prospective cohorts, manual OBP and ABP measurements were compared among ambulatory hypertensive patients. Hypertension was defined as systolic OBP ≥ 140 mmHg and/or diastolic OBP ≥ 90 mmHg, and systolic ABP ≥ 130 mmHg and/or diastolic ABP ≥ 80 mmHg.
RESULTS
In the retrospective cohort (n = 903), the mean OBP1 (before ABP measurement) was higher than ABP in both systolic (138 ± 17 mmHg vs. 123 ± 13 mmHg, p < 0.001) and diastolic (84 ± 12 mmHg vs. 78 ± 11 mmHg, p < 0.001) measurements. Interestingly, there was only a weak correlation between OBP and ABP (r² = 0.038, p < 0.001). The overall discordance rate of OBP compared to ABP, which is the reference method for measuring BP, was 43.9%. The prospective cohort (n = 57) showed similar results. In a subgroup analysis, male patients had higher false negative results (masked or under-treated hypertension) than did female patients (26.1% vs. 17.8%, p = 0.003), whereas female patients had a higher false positive rate (white-coat or over-treated hypertension) than did male patients (28.7% vs. 15.2%, p < 0.001).
CONCLUSIONS
The diagnostic accuracy of manual OBP is low in reference to ABP. Men and women have different patterns of discordance. These findings indicate that management of hypertensive patients with manual OBP measurements may be suboptimal and encourages the use of ABP in ambulatory hypertensive patients.

Keyword

Manual office blood pressure; Ambulatory blood pressure; Diagnostic accuracy; Sex

MeSH Terms

Blood Pressure*
Cohort Studies
Female
Humans
Hypertension
Korea*
Male
Methods
Prospective Studies
Retrospective Studies
Full Text Links
  • KJIM
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr