J Korean Radiol Soc.  1996 Feb;34(2):223-229.

Reduced Volume of Contrast Media: Effect on Vascular Opacification and Image Quality in Spiral CT of the Chest

Affiliations
  • 1Department of Diagnostic Radiology, Asan Medical Center, University of Ulsan College of Medicine, Korea.

Abstract

PURPOSE
To evaluate the effect of reduced volume of contrast media on vascular opacification and image quality in spiral CT of the chest.
MATERIALS AND METHODS
Sixty patients referred for chest CT were examined withspiral CT with 60ml(n=30) or 90ml(n=30) of 30% ionic contrast media(Rayvist 300 , Schering, Germany) alternately.Injection rate of each group was as follows : 2.0 ml/sec for 20 seconds followed by 1.0 ml/sec for 20 seconds in 60ml group and 2.0ml/sec for 45 seconds in 90ml group. Twenty-five seconds scanning delay was employed. For the objective comparison of vascular opacification, CT numbers were measured at superior vena cava, ascending and descending aorta, right and left pulmonary artery, left atrium, and inferior vena cava. For the subjective comparison three radiologists scored the grade of vascular opacification and image quality blindly and independently. All data were analyzed statistically.
RESULTS
The mean values of measured CT numbers in 90ml group were higher than those in 60ml group(p <.05) at the same level. The overall mean score of vascular opacification in 90ml group was 2.86, and 2.31 in 60ml group(p <.0001). In the overall mean score of imagequality, there was no statistically significant difference between 90ml group(2.46) and 60ml group(2.40).
CONCLUSION
Althought there is some degradation of vascular opacification in 60ml group, overall image quality is not degraded. Therefore, 60 ml of contrast media can be used in spiral CT of the chest without degradation ofoverall image quality except in spiral CT angiography.

Keyword

Computed tomography(CT), contrast media; Computed tomography(CT), helical; Constrast media, comparative studies; Contrast media, effects; Thorax, CT

MeSH Terms

Angiography
Aorta
Contrast Media*
Heart Atria
Humans
Pulmonary Artery
Thorax*
Tomography, Spiral Computed*
Tomography, X-Ray Computed
Vena Cava, Inferior
Vena Cava, Superior
Contrast Media
Full Text Links
  • JKRS
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr