J Korean Ophthalmol Soc.  2002 Jun;43(6):1046-1050.

Change of intraocular pressure, corneal flap and thickness by automated corneal shaper(R) in porcine eye

Affiliations
  • 1Department of Ophthalmology, Yonsei University College of Medicine, Seoul, Korea. jblee88@yumc.yonsei.ac.kr
  • 2Institute of Vision Research, Yonsei University College of Medicine, Seoul, Korea.

Abstract

PURPOSE: To evaluate the performance of Automated Corneal Shaper(R)(ACS)
METHODS
we analysed the changes in intraocular pressure, axial length and corneal thickness and its influence on corneal flap following keratectomy by ACS in 20 porcine eyes.
RESULTS
Corneal flap was made by ACS with plate 160, and its size and thickness were measured. With the suction ring application, intraocular pressure rised from 17.0 +/-1.96 mmHg to 62.0 +/-11.8 mmHg (p=0.001), while corneal thickness and axial length were decreased from 1090 +/-75.1 micro meter to 1074 +/-73.7 micro meter(p=0.001) and from 19.94 +/-0.89 to 19.72 +/-0.73 mm (p=0.028) respectively. Thickness of corneal flap was 122 +/-23.9 micro meter and its diameter was 8.3 +/-0.15 mm in average.
CONCLUSION
special attention is required during LASIK (laser in situ keratomileusis) with Automated Corneal Shaper(R) in order to prevent intraocular damage by abrupt rise in intraocular pressure and change of eyeball contour.

Keyword

Automated microkeratome; Corneal flap; Intraocular damage; LASIK; Suction ring

MeSH Terms

Intraocular Pressure*
Keratomileusis, Laser In Situ
Suction
Full Text Links
  • JKOS
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr