Prog Med Phys.  2016 Jun;27(2):79-85. 10.14316/pmp.2016.27.2.79.

The Dose Attenuation according to the Gantry Angle and the Photon Energy Using the Standard Exact Couch and the 6D Robotic Couch

Affiliations
  • 1Department of Physics, Yeungnam University, Gyeongsan, Korea.
  • 2Department of Radiation Oncology, Yeungnam University Medical Center, Daegu, Korea.
  • 3Department of Radiation Oncology, Yeungnam University College of Medicine, Daegu, Korea. skkim3@ynu.ac.kr

Abstract

The objective of this study is to increase the accuracy of dose transmission in radiation therapy using two types of treatment tables, standard exact couch (Varian 21EX, Varian Medical Systems, Milpitas, CA) and 6D robotic couch (Novalis, BrainLAB A.G., Heimstetten, Germany)). We examined the dose attenuation based on the two types of treatment tables and studied the dose of attenuation using the phase (In/Out) for the standard exact couch. We measured the relative dose according to the incident angle of a penetrative photon beam under a treatment table. The incident angle of the photon beam was from 0° to 360° in the increments of 5°. The reference angle was set to 0°. Furthermore, the relative dose of the 6D robotic couch was measured using 6 MV and 15 MV, and that of the standard exact couch was measured at the sliding rail position (In-Out) using 6 MV and 10 MV. In the case of the standard exact couch, the measured relative dose was 16.53% (rails at the "In position," 175°, 6 MV), 12.42% (rails at the "In position," 175°, 10 MV), 13.13% (rails at the "Out position," 175°, 6 MV), and 9.96% (rails at the "Out position," 175°, 10 MV). In the case of the 6D robotic couch, the measured relative dose was 6.82% (130°, 6 MV) and 4.92% (130°, 15 MV). The photon energies were surveyed at the same incident angle. The dose attenuation for an energy of 10 MV was 4~5% lower than that for 6 MV. This indicated that the higher photon energy, lesser is the attenuation. The results of this study indicated that the attenuation rate for the 6D robotic couch was confirmed to be 1% larger than that for the standard exact couch at 6 MV and 180°. In the case of the standard exact couch, the dose attenuation was found to change rapidly in accordance with the phase ("In position" and "Out position") of the sliding rail.

Keyword

Treatment couch; Status; Dose; Incident angle; Attenuation rate

Figure

  • Fig. 1. The photographs of treatment couch: (a) Standard exact couch (Varian Medical Systems, Milpitas, CA), (b) 6D robotic couch (BrainLAB A.G., Heimstetten, Germany). (a-1) The side view of the standard exact couch, (a-2) The back side view of the standard exact couch, (a-3) The top view of the standard exact couch. (b-1) The side view of the 6D robotic couch, (b-2) The back side view of the 6D robotic couch, (b-3) The top view of the 6D robotic couch.

  • Fig. 2. The photographs of universal sandwich panel rail position: (a) Bottom view of rail at the “In position” and upper carbon fiber layer apart. (b) Bottom view of rail at the “Out position” and upper carbon fiber layer apart.

  • Fig. 3. Set up images for measurement of transmission and attenuation rate from a cylinder acrylic phantom.

  • Fig. 4. The schematic diagram of the dose measurement using the ion chamber and cylinder acrylic phantom. 18 cm diameter cylindrical phantom centered on couch with ion chamber at center, beams every 5o from 0o to 360o. θ means the included angle from the couch's vertical axis to the gantry's incident angle.

  • Fig. 5. Relative dose of the standard exact couch's rail at the “in position”, field size 5×5 cm.

  • Fig. 6. Relative dose of the standard exact couch's rail at the “out position”, field size 5×5 cm.

  • Fig. 7. Relative dose of the 6D robotic couch, field size 5×5 cm.

  • Fig. 8. Dependence according to photon beam's incident angle and the photon energy about the attenuation rate for the type of the couch.


Reference

References

1. Oh SA, Kang MK, Kim SK. Comparison of IMRT and VMAT Techniques in Spine Stereotactic Radiosurgery with International Spine Radiosurgery Consortium Consensus Guidelines. Prog Med Phys 24, 145–153 (. 2013.
2. Donovana E, Bleakleya N, Denholmb E, Evansa P, et al. Randomised trial of standard 2D radiotherapy (RT) versus intensity modulated radiotherapy (IMRT) in patients prescribed breast radiotherapy. Radiotherapy and Oncology 82, 254–264 (. 2007.
3. Ahn YC.: Introduction of intensity modulated radiation therapy. J Korean Med Assoc 54 (. 2011.
4. Otto K. Volumetric modulated arc therapy: IMRT in a single gantry arc. Medcal physics 35 (. 2008.
5. Uematsu M, Shioda A, Tahara K, et al. Focal, High Dose, and Fractionated Modified Stereotactic Radiation Therapy for Lung Carcinoma Patients. American Cancer Society 82, 1062–1070 (. 2000.
6. Papanikolaou N, Battista JJ, Boyer AL, et al. Task Group 65 report: Tissue inhomogeneity corrections for megavoltage photon beams. (. 2004.
7. Kleinb EE, Hanley J, Bayouth J, et al. Task Group 142 report: Quality assurance of medical accelerators. (. 2009.
8. Han SC, Lee DJ.: Differences in Target Volume Delineation Using Typical Radiosurgery Planning System. PROGRESS in MEDICAL PHYSICS 24, 265–270 (. 2013.
9. Olch AJ, Gerig L, Li H, et al. Dosimetric effects caused by couch tops and immobilization devices: report of AAPM Task Group 176. Medical physics 41, 061501,. doi: 10.1118/1.4876299. 2014.
10. Johns H, Cunningham JR. Physics of Radiology 4th. 137–142 (. 1983.
11. Wagner D, Vorwerk H.: Treatment couch modeling in the treatment planning system Eclipse J. Cancer Sci. Ther 3 007–012 (. 2011.
12. Van Prooijen M, Kanesalingam T, Islam MK. Assessment and management of radiotherapy beam intersections with the treatment couch. JOURN AL OF APPLIED CLINICAL MEDICAL PHYSICS 11, 128–139 (. 2010.
13. Kim SJ, Sin HS, Ko SY, et al. A Correction Method of Dose to Attenuation Rate of Transmitting Photon Beam Through Couch Top for Radiosurgery Using Novalis. PROGRESS in MEDICAL PHYSICS 22, 12–17 (. 2011.
14. Li H, Lee AK, Johnson JL, et al. Characterization of dose impact on IMRT and VMAT from couch attenuation for two Varian couches. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS 12, 23–31 (. 2011.
15. Njeh CF, Raines TW, Saunders MW. Determination of the photon beam attenuation by the BrainLAB imaging couch angular and field size dependence. JOURN AL OF APPLIED CLINICAL MEDICAL PHYSICS 10, 16–27 (. 2009.
16. Spezi E, Ferri A. Dosimetric characteristics of the Siemens IGRT carbon fiber tabletop. Medical dosimetry: official journal of the American Association of Medical Dosimetrists 32, 295-298,. doi: 10.1016/j.meddos.2006.11.006. 2007.
17. Brooks RA, Chiro GD. Beam Hardening in X-ray Rec.
Full Text Links
  • PMP
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr