Korean J Med Phys.  2011 Mar;22(1):12-17.

A Correction Method of Dose to Attenuation Rate of Transmitting Photon Beam Through Couch Top for Radiosurgery Using Novalis

Affiliations
  • 1Department of Physics, College of Science, Yeungnam University, Gyongsan, Korea.
  • 2Department of Therapeutic Radiology & Oncology, CHA Bundang Medical Center, CHA University, Seongnam, Korea.
  • 3Department of Therapeutic Radiology & Oncology, College of Medicine, Yeungnam University, Daegu, Korea. skkim@med.yu.ac.kr

Abstract

This study has its own goal to deliver the accurate dose on the target volume by calculating and modifying the attenuation rate of photon beam transmitting the couch top with geometric model. The experiment was that the transmission rate and attenuation rate of photon beam transmitting the couch top was predicted by the geometric model, then compared and analyzed with what was measured experimentally based on that. The result showed that the predicted value by the geometric model accorded closely with the experimental value. In addition, in order to judge whether the practical clinical application is available, the point dose, measured after modifying the attenuation rate modelinged according to the treatment plan of a patient of spine radiosurgery, was compared with the one done nothing. The result was that the former showed decreased error range with treatment planned one than the latter. This papers calculated the transmission and attenuation rate with the geometric model transmitting the couch top and verified it experimentally. This method is expected to be very useful in not only the radiosurgery using Novalis but also the general radiation therapy.

Keyword

Novalis; Couch top; Radiosurgery; Transmission rate; Attenuation rate

MeSH Terms

Humans
Radiosurgery
Spine
Full Text Links
  • KJMP
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr