J Korean Soc Ther Radiol.  1991 Dec;9(2):337-342.

The Comparison of the KAPM Dosimetric Protocol (1990) with the TG-21 and Clambda/C(E) Method

Affiliations
  • 1Department of Therapeutic Radiology, Asan Medical Center, College of Medicine, University of Ulsan, Seoul, Korea.

Abstract

The Comparison of the KAPM Dosimetric Protocol (1990) with the TG-21 and Clambda/C(E) (ICRU-21 and SCRAD protocol) method is studied. The therapetutic range of radiation (photon 4MV-15MV and electron 6MeV-20MeV)and three kinds of the chambers were used in the water phantom. The Results from TG-21 and KAPM Protocol did not show much differences (less than 1%) throughout the whole energy range; N(D) from KAPM protocol and Ngas from TG-21 showed 0.2% deviation mainly from W/e difference between two protocols. But the results from KAPM protocol (1990) and those from Clambda/C(E) Method showed -1.9+/-0.6%(KAPM protocol is higher) deviation for photom beam and +3.3+/-1% (KAPM protocol is lower) deviation for electron beams.

Keyword

Dosimetric Protocol; KAPM Dosimetric Protocol; TG-21

MeSH Terms