Restor Dent Endod.  2012 May;37(2):68-73.

An evaluation of the accuracy of Root ZX according to the conditions of major apical foramen

Affiliations
  • 1Department of Conservative Dentistry, Chosun University School of Dentistry, Gwangju, Korea. rootcanal@hanmail.net

Abstract


OBJECTIVES
The purpose of this study was to assess the accuracy of Root ZX (J. Morita Corp.) according to the location of major foramen and open apex.
MATERIALS AND METHODS
81 mandibular premolars with mature apices were selected. After access preparation, 27 teeth were instrumented to simulate open apices. 54 teeth were classified according to location of major foramen under surgical microscope (x16). The file was fixed at the location of apical constriction by Root ZX using glass ionomer cement. The apical 4 mm of the apex was exposed and photo was taken and the distance from file tip to the major foramen was measured by calibrating metal ruler on graph paper. The results were statistically analyzed using ANOVA and Scheffe test at p < 0.05 level.
RESULTS
Mean distance from file tip to major foramen was 0.308 mm in Tip foramen group (I), 0.519 mm in Lateral foramen group (II) and 0.932 mm in open apex group (III). Root ZX located apical constriction accurately within +/- 0.5 mm in group I of 85.71%, in group II of 59.09%, and in group III of 33.33%. There was a statistically significant difference between group I and III (p < 0.05).
CONCLUSION
Root ZX located apical constriction accurately regardless of location of major foramen. However, Root ZX couldn't find it in open apex. Clinicians have to use a combination of methods to determine an appropriate working length at open apex. It may be more successful than relying on just electronic apex locator.

Keyword

Accuracy; Electronic apex locator; Major foramen; Open apex

MeSH Terms

Acrylic Resins
Bicuspid
Constriction
Electronics
Electrons
Glass Ionomer Cements
Silicon Dioxide
Tooth
Tooth Apex
Acrylic Resins
Glass Ionomer Cements
Silicon Dioxide

Figure

  • Figure 1 An open apex model. (a) The canals were instrumented with a #2 Peeso reamer to the actual length; (b) A divergent open apex was prepared by retrograde apical preparation with a .04/50 K3 file inserted to the length of the cutting blade; (c) Simulated open apex model.

  • Figure 2 Tip foramen (group I), in which major foramen is located at the tip along the main axis of root. File was fixed in the position and the distance from file tip to major foramen was measured (×25).

  • Figure 3 Lateral foramen (group II), in which major foramen deviates from the main axis of root. File was fixed in the position and the distance from file tip to major foramen was measured (×25).

  • Figure 4 Simulated open apex (group III). File was fixed in the position and the distance from file tip to major foramen was measured (×25).


Reference

1. Ding J, Gutmann JL, Fan B, Lu Y, Chen H. Investigation of apex locators and related morphological factors. J Endod. 2010. 36:1399–1403.
Article
2. Ricucci D, Langeland K. Apical limit of root canal instrumentation and obturation, part 2. a histological study. Int Endod J. 1998. 31:394–409.
Article
3. Suzuki K. Experimental study on iontophoresis. J Jpn Stomatol. 1942. 16:411–417.
4. Sunada I. New method for measuring the length of the root canal. J Dent Res. 1962. 41:375–387.
Article
5. Kobayashi C, Suda H. New electronic canal measuring device based on the ratio method. J Endod. 1994. 20:111–114.
Article
6. Kobayashi C. Electronic canal length measurement. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 1995. 79:226–231.
Article
7. Green D. Stereomicroscopic study of 700 root apices of maxillary and mandibular posterior teeth. Oral Surg Oral Med Oral Pathol. 1960. 13:728–733.
Article
8. Dummer PM, McGinn JH, Rees DG. The position and topography of the apical canal constriction and apical foramen. Int Endod J. 1984. 17:192–198.
Article
9. Kuttler Y. Microscopic investigation of root apexes. J Am Dent Assoc. 1955. 50:544–552.
Article
10. Torabinejad M, Walton RE. Endodontics: principles and practice. 2009. 4th ed. Missouri: Saunders Elsevier;29–34.
11. Herrera M, Abalos C, Planas AJ, Llamas R. Influence of apical constriction diameter on Root ZX apex locator precision. J Endod. 2007. 33:995–998.
Article
12. Stein TJ, Corcoran JF, Zillich RM. Influence of the major and minor foramen diameters on apical electronic probe measurements. J Endod. 1990. 16:520–522.
13. Fouad AF, Rivera EM, Krell KV. Accuracy of the Endex with variations in canal irrigants and foramen size. J Endod. 1993. 19:63–67.
Article
14. Huang L. An experimental study of the principle of electronic root canal measurement. J Endod. 1987. 13:60–64.
Article
15. Hachmeister DR, Schindler WG, Walker WA 3rd, Thomas DD. The sealing ability and retention characteristics of mineral trioxide aggregate in a model of apexification. J Endod. 2002. 28:386–390.
Article
16. Cho JH, Kum KY, Lee SJ. In vitro evaluation of accuracy and consistency of four different electronic apex locators. J Korean Acad Conserv Dent. 2006. 31:390–397.
Article
17. Ponce EH, Vilar Fernández JA. The cemento-dentino-canal junction, the apical foramen, and the apical constriction: evaluation by optical microscopy. J Endod. 2003. 29:214–219.
Article
18. Sjogren U, Hägglund B, Sundqvist G, Wing K. Factors affecting the long-term results of endodontic treatment. J Endod. 1990. 16:498–504.
Article
19. Pagavino G, Pace R, Baccetti T. A SEM study of in vivo accuracy of the Root ZX electronic apex locator. J Endod. 1998. 24:438–441.
20. Goldberg F, De Silvio AC, Manfré S, Nastri N. In vitro measurement accuracy of an electronic apex locator in teeth with simulated apical root resorption. J Endod. 2002. 28:461–463.
Article
21. Nguyen HQ, Kaufman AY, Komorowski RC, Friedman S. Electronic length measurement using small and large files in enlarged canals. Int Endod J. 1996. 29:359–364.
Article
22. Kim BH, Lee YK, Kim YS. A study on the accuracy of the ROOT-ZX in the canal with mechanically formed constriction. J Korean Acad Conserv Dent. 1999. 24:628–632.
23. Czerw RJ, Fulkerson MS, Donnelly JC. An in vitro test of a simplified model to demonstrate the operation of electronic root canal measuring devices. J Endod. 1994. 20:605–606.
Article
24. Cho YL, Son WH, Hwang HK. An accuracy of the several electronic apex locators on the mesial root canal of the mandibular molar. J Korean Acad Conserv Dent. 2005. 30:477–485.
Article
25. Lee SJ, Nam KC, Kim YJ, Kim DW. Clinical accuracy of a new apex locator with an automatic compensation circuit. J Endod. 2002. 28:706–709.
Article
26. Ounsi HF, Naaman A. In vitro evaluation of the reliability of the Root ZX electronic apex locator. Int Endod J. 1999. 32:120–123.
Article
27. Mayeda DL, Simon JH, Aimar DF, Finley K. In vivo measurement accuracy in vital and necrotic canals with the Endex apex locator. J Endod. 1993. 19:545–548.
Article
28. Herrera M, Abalos C, Planas AJ, Llamas R. Influence of apical constriction diameter on Root ZX apex locator precision. J Endod. 2007. 33:995–998.
Article
29. Shabahang S, Goon WW, Gluskin AH. An in vivo evaluation of Root ZX electronic apex locator. J Endod. 1996. 22:616–618.
30. Tselnik M, Baumgartner JC, Marshall JG. An evaluation of root ZX and elements diagnostic apex locators. J Endod. 2005. 31:507–509.
Article
31. Meredith N, Gulabivala K. Electrical impedance measurements of root canal length. Endod Dent Traumatol. 1997. 13:126–131.
Article
32. Nekoofar MH, Ghandi MM, Hayes SJ, Dummer PM. The fundamental operating principles of electronic root canal length measurement devices. Int Endod J. 2006. 39:595–609.
Article
33. Berman LH, Fleischman SB. Evaluation of the accuracy of the Neosono-D electronic apex locator. J Endod. 1984. 10:164–167.
Article
34. Saito T, Yamashita Y. Electronic determination of root canal length by newly developed measuring device. Influences of the diameter of apical foramen, the size of K-file and the root canal irrigants. Dent Jpn (Tokyo). 1990. 27:65–72.
35. Huang L. An experimental study of the principle of electronic root canal measurement. J Endod. 1987. 13:60–64.
Article
36. ElAyouti A, Weiger R, Löst C. Frequency of overinstrumentation with an acceptable radiographic working length. J Endod. 2001. 27:49–52.
Article
Full Text Links
  • RDE
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr