J Korean Acad Conserv Dent.  2006 Sep;31(5):390-397. 10.5395/JKACD.2006.31.5.390.

In vitro evaluation of accuracy and consistency of four different electronic apex locators

Affiliations
  • 1Department of Conservative Dentistry, Oral Science Research Center, Yonsei University, Seoul, Korea. sjlee@yumc.yonsei.ac.kr
  • 2Department of Conservative Dentistry, Dental Research Institute, Seoul National University, Seoul, Korea.

Abstract

The purpose of this study was to evaluate the accuracy and the consistency of four different electronic apex locators in an in vitro model. Fourty extracted premolars were used for the study. Four electronic apex locators (EAL) were Root ZX, SmarPex, Elements Diagnostic Unit (EDU), and E-Magic Finder Deluxe (EMF). After access preparation, the teeth were embedded in an alginate model and the length measurements were carried out at "0.5"and "Apex"mark using four EALs. The file was cemented at the location of the manufacturers'instruction (Root ZX, EDU, EMF: 0.5 mark, SmarPex: Apex mark). The apical 4mm of the apex was exposed and the distance from the file tip to the major foramen was measured by Image ProPlus (x 100). The distance from the file tip to the major foramen was calculated at 0.5 and Apex mark and the consistency of 0.5 and Apex mark was compared by SD and Quartile of Box plots. In this study, Root ZX and EMF located the apical constriction accurately within +/- 0.5 mm in 100%, whereas SmarPex and EDU located in 90% and in 70% respectively. For Root ZX and EMF, there was no significant difference between the consistency of 0.5 and Apex mark. However, for the EDU and SmarPex, Apex mark was more consistent than 0.5 mark. From the evaluation of the consistency in this study, for Root ZX and EMF, both 0.5 and Apex mark can be used as a standard mark. And for EDU and SmarPex, the Apex mark can be recommended to be used as a standard mark.

Keyword

Electronic apex locator; Consistency; Accuracy; Alginate model; Major foramen; Apical constriction

MeSH Terms

Bicuspid
Constriction
Tooth

Figure

  • Figure 1 Alginate model.

  • Figure 2 Length measurement with Digital caliper.

  • Figure 3 Distance measurement between the file tip and major foramen by Image ProPlus.

  • Figure 4 Box plots : Distance between the file tip and the major foramen at 0.5 and Apex mark (Quartile: the value of 75%-the value of 25%, The mark in which Quartile is smaller is more consistent).


Cited by  1 articles

An evaluation of the accuracy of Root ZX according to the conditions of major apical foramen
Shin-Young Park, Dong-Kyun Lee, Ho-Keel Hwang
Restor Dent Endod. 2012;37(2):68-73.    doi: 10.5395/rde.2012.37.2.68.


Reference

1. Grove C. Why canals should be filled to the dentinocemental junction. J Am Dent Assoc. 1930. 17:293–296.
2. Ricucci D, Langeland K. Apical limit of root canal instrumentation and obturation. Int Endod J. 1998. 31:394–409.
3. Suzuki K. Experimental study on iontophoresis. Jpn J Stomatol. 1942. 16:411–429.
4. Sunada I. New method for the measuring the length of the root canal. J Dent Res. 1962. 41:375–387.
Article
5. Ushiyama J. New Principle and method for measuring the root canal length. J Endod. 1983. 9:97–104.
Article
6. Hasegawa K, Iizuka H, Takei M, Goto N, Nihei M, Ohashi M. A new method and apparatus for measuring root canal length. J Nihon Univ Sch Dent. 1986. 28:117–128.
Article
7. Frank AL, Torabinejad M. An in vivo evaluation of Endex electronic apex locator. J Endod. 1993. 19:177–179.
8. Kobayashi C, Okiji T, Kaqwashima N, Suda H, Sunada I. A basic study on the electronic root canal length measurement: Part 3. Newly designed electronic root canal length measuring device using division method. Jpn J Conserv Dent. 1991. 34:1442–1448.
9. Kobayashi C, Suda H. New electronic canal measuring device based on the ratio method. J Endod. 1994. 20:111–114.
Article
10. Kobayashi C. Electronic canal length measurement. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 1995. 79:226–231.
Article
11. Shabahang S, Goon WW, Gluskin AH. An in vivo evaluation of Root ZX electronic apex locator. J Endod. 1996. 22:616–618.
12. Dunlap CA, Remeikis NA, BeGole EA, Rauschenberger CR. An in vivo evaluation of an electronic apex locator that uses the ratio method. J Endod. 1998. 24:48–50.
Article
13. Welk AR, Baumgartner JC, Marshall JG. An in vivo comparison of two frequency-based electronic apex locator. J Endod. 2003. 29:497–500.
14. Meares WA, Steiman HR. The influence of sodium hypochlorite irrigation on the accuracy of the Root ZX electronic apex locator. J Endod. 2002. 28:595–598.
Article
15. Jenkins JA, Walker WA 3rd, Schindler WG, Flores CM. An in vitro evaluation of the accuracy of the Root ZX in the presence of various irrigants. J Endod. 2001. 27:209–211.
Article
16. Lee SJ, Nam KC, Kim YJ, Kim DW. Clinical accuracy of a new apex locator with an automatic compensation circuit. J Endod. 2002. 28:706–709.
Article
17. Kaufman AY, Keila S, Yoshpe M. Accuracy of a new apex locator: an in vitro study. Int Endod J. 2002. 35:186–192.
18. Kuttler Y. Microscopic investigation of root apexes. J Am Dent Assoc. 1955. 50:544–552.
Article
19. Dummer PM, McGinn JH, Rees DG. The position and topography of the apical canal constriction and apical foramen. Int Endod J. 1984. 17:192–198.
Article
20. Mayeda DL, Simon JH, Aimar DF, Finley K. In vivo measurement accuracy in vital and necrotic canals with the Endex apex locator. J Endod. 1993. 19:545–548.
Article
Full Text Links
  • JKACD
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr