J Korean Acad Conserv Dent.  2006 Sep;31(5):371-377. 10.5395/JKACD.2006.31.5.371.

In vitro comparison of measurement accuracy in pre-enlarged and enlarged canals with four apex locators

Affiliations
  • 1Department of Conservative Dentistry, School of Dentistry, Pusan National University, Busan, Korea. golddent@pusan.ac.kr

Abstract

The purposes of this study were to assess the accuracy of measurements in pre-enlarged canals with small instruments and to compare the accuracies, in enlarged canal, with small size instruments and instruments that match the actual canal diameter using Root ZX, Bingo1020, SmarPex, and e-Magic Finder. Ten extracted teeth were embedded in an alginate model made for testing apex locators. A size 10 file was placed into the root canal until the tip of the file reached the plane of the major diameter of the foramen under a dental operating microscope at the 25 x magnification. The measurement was done with digital caliper and defined as actual length. Electronic length measurement with a size 10 file in pre-enlarged canal was done by reading the index indicating Apex of each device to gain a definite value. After completion of canal enlargement to a size 45 file, each difference between actual length and electric measurement value with a size 10 and 40 files in enlarged canal was recorded as L10 and L40. The one-way ANOVA and Scheffe's multiple range tests were computed for analyze the differences among the four apex locators in the same group. The Student's t-test between L10 and L40 of each locator was done. The accuracies of electronic measurements were significantly different among the 4 devices. The file size made no difference on the accuracy of electronic measurement in enlarged canal with same device. The e-Magic Finder was the most accurate device among the 4 apex locators used in this study.

Keyword

Electronic Apex Locator; File size; Measurement accuracy; Working length

MeSH Terms

Dental Pulp Cavity
Tooth

Figure

  • Figure 1 Schematic diagram of alginate model (a: file, b: lip clip, c: apex locator, d: alginate).


Reference

1. Sjögren U, Haggluland B, Sunquist G, Wing K. Factors affecting the long-term results of endodontic treatment. J Endod. 1990. 16:498–504.
Article
2. Kaufman AY, Katz A. Reliability of Root ZX apex locator tested by an in vitro model. J Endod. 1993. 19:201. (abstract).
3. Dunlap CA, Remeikis NA, BeGole EA, Rauschenberger CR. An in vivo evaluation of an electronic apex locator that uses the ratio method in vital and necrotic canals. J Endod. 1998. 24:48–50.
Article
4. Sunada I. New method for measuring the length of the root canal. J Dent Res. 1962. 41:375–387.
Article
5. McDonald NJ, Hovland EJ. An evaluation of the Apex Locator Endocater. J Endod. 1990. 16:5–8.
Article
6. O'Neill LJ. A clinical evaluation of electronic root canal measurement. Oral Surg Oral Med Oral Pathol. 1974. 38:469–473.
7. Trope M, Rabie G, Tronstad L. Accuracy of an electronic apex locator under controlled clinical conditions. Endod Dent Traumatol. 1985. 1:142–145.
Article
8. Kobayashi C, Suda H. New electronic canal measuring device based on the ratio method. J Endod. 1994. 20:111–114.
Article
9. Kobayashi C. Electronic canal length measurement. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 1995. 79:226–231.
Article
10. Fouad AF, Rivera EM, Krell KV. Accuracy of the Endex with variations in canal irrigants and foramen size. J Endod. 1993. 19:63–67.
Article
11. Frank AL, Torabinejad M. An in vivo evaluation of Endex electronic apex locator. J Endod. 1993. 19:177–179.
12. Mayeda DL, Simon JH, Aimar DF, Finley K. In vivo measurement accuracy in vital and necrotic canals with the Endex apex locator. J Endod. 1993. 19:545–548.
Article
13. Fouad AF, Krell KV. An in vitro comparison of five root canal length measuring instruments. J Endod. 1989. 15:573–577.
14. Operating Instructions of Root ZX. 1998. Kyoto, Japan: J. Morita MFG. Corp..
15. Kuttler Y. Microscopic investigation of root apexes. J Am Dent Assoc. 1955. 50:544–552.
Article
16. Tinaz AC, Alacam T, Topuz Ö. A simple model to demonstrate the electronic apex locator. Int Endod J. 2002. 35:940–945.
Article
17. Kaufman AY, Keila S, Yoshpe M. Accuracy of a new apex locator: an in vitro study. Int Endod J. 2002. 35:186–192.
18. Ounsi HF, Naaman A. In vitro evaluation of the realiability of the Root ZX electronic apex locator. Int Endod J. 1999. 32:120–123.
Article
19. Nguyen HQ, Kaufman AY, Komowski RC, Friedman S. Electronic length measuring using small and large files in enlarged canals. Int Endod J. 1996. 29:359–364.
Article
20. Shabahang S, Goon WWY, Gluskin AH. An in vivo evaluation of Root ZX electronic apex locator. J Endod. 1996. 22:616–618.
21. Welk AR, Baumgartner JC, Marshall JG. An in vivo comparison of two frequency-based electronic apex locators. J Endod. 2003. 29:497–500.
22. Donnelly JC. A simplified model to demonstrate the operation of electronic root canal measuring devices. J Endod. 1993. 19:579–580.
Article
23. Czerw RJ, Fulkerson MS, Donnelly JC. An in vitro test of a simplified model to demonstrate the operation of electronic root canal measuring devices. J Endod. 1994. 20:605–606.
Article
24. Fouad AF, Krell KV, McKendry DJ, Koorbusch GF, Olson RA. A clinical evaluation of five electronic root canal length measuring instruments. J Endod. 1990. 16:446–449.
Article
25. Ricard O, Roux D, Bourdeau L, Woda A. Clinical evaluation of the accuracy of the evident RCM Mark II apex locator. J Endod. 1991. 17:567–569.
Article
Full Text Links
  • JKACD
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr