Radiat Oncol J.
2012 Sep;30(3):146-151.
The impact of beam angle configuration of intensity-modulated radiotherapy in the hepatocellular carcinoma
- Affiliations
-
- 1Department of Radiation Oncology, Yeungnam University College of Medicine, Daegu, Korea. mkkang@ynu.ac.kr
- 2Department of Physics, Yeungnam University, Gyeongsan, Korea.
Abstract
- PURPOSE
This treatment planning study was undertaken to evaluate the impact of beam angle configuration of intensity-modulated radiotherapy (IMRT) on the dose of the normal liver in hepatocellular carcinoma (HCC).
MATERIALS AND METHODS
The computed tomography datasets of 25 patients treated with IMRT for HCC were selected. Two IMRT plans using five beams were made in each patient; beams with equidistance of 72degrees (Plan I), and beams with a 30degrees angle of separation entering the body near the tumor (Plan II). Both plans were generated using the same constraints in each patient. Conformity index (CI), homogeneity index (HI), gamma index, mean dose of the normal liver (Dmean_NL), Dmean_NL difference between the two plans, and percentage normal liver volumes receiving at least 10, 20, and 30 Gy (V10, V20, and V30) were evaluated and compared.
RESULTS
Dmean_NL, V10, and V20 were significantly better for Plan II. The Dmean_NL was significantly lower for peripheral (p = 0.001) and central tumors (p = 0.034). Dmean_NL differences between the two plans increased in proportion to gross tumor volume to normal liver volume ratios (p = 0.002). CI, HI, and gamma indices were not significantly different for the two plans.
CONCLUSION
The IMRT plan based on beams with narrow separations reduced the irradiated dose of the normal liver, which would allow radiation dose escalation for HCC.