Korean J Med Phys.  2007 Sep;18(3):139-143.

Dose Measurements using Phantoms for Tube Voltage, Tube Current, Slice Thickness in MDCT

Affiliations
  • 1Department of Radiological Science, College of Health Science and Research Institute of Health Science, Yonsei University. hjk1@yonsei.ac.kr
  • 2Department of Radiology, Wonju Christian Hospital.

Abstract

The purpose of this study was to measure and evaluate radiation dose for MDCT parameters. Patient dose for various combination of MDCT parameters were experimentally measured, using MDCT (GE light speed plus 4 slice, USA), model 2026C electrometer (RADICAL 2026C, USA), standard Polymethylmethacrylate (PMMA) head and body CT dosimetry phantoms. In clinical situations, for a typical abdominal scan performed with MDCT at 120 kVp, 180 mAs, 20 mm collimation, and a pitch of 0.75, CTDIw, CTDIvol were measured as 20.2 mGy, 26.9 mGy, respectively. When scan length is assumed as 271.3 mm, DLP and measured effective dose of the abdominal would be calculated as 729.1 mGy cm, 10.9 mSv, respectively.

Keyword

MDCT; Tube voltage; Tube current; Beam collimation; CTDIw

MeSH Terms

Head
Humans
Polymethyl Methacrylate
Polymethyl Methacrylate
Full Text Links
  • KJMP
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr