Korean J Lab Med.  2006 Aug;26(4):294-298. 10.3343/kjlm.2006.26.4.294.

Networking Experience of Point-of-Care Test Glucometer

Affiliations
  • 1Department of Laboratory Medicine, Chung-Ang University College of Medicine, Seoul, Korea. ajcp@unitel.co.kr

Abstract

BACKGROUND: Clinical laboratories are selecting total solutions to point-of-care test (POCT) by handling all aspects of information management ranging from the ordering of the test to the reporting of verified results. Today, many POCT devices can be directly connected to the laboratory information system (LIS) via data management system (DMS). We report here our experience with the networking of POCT glucometers.
METHODS
We set up computerization of POCT glucometer (Precision PCx, Abbott, USA) in August, 2004 and established a quality control (QC) system for the portable glucometers on 16 wards in Chung-Ang University Hospital. Precision of the glucometers was assessed by performing 28 replicate tests using quality control specimens containing 5 different glucose concentrations. Turnaround time (TAT) was calculated from the data of 17,796 glucose tests performed during a 5-month period in 2004 before computerization of the glucometers and 31,513 tests performed during the same 5-month period of 2005 after computerization.
RESULTS
In the precision study, within-run and total-run CVs were within 10%. With the networking of POCT glucometers, the TAT from the collection of specimen to the specimen receipt to the LIS was 28.5+/-17.2 min compared to 43.6+/-17.5 without the networking (P<0.0001). A staff satisfaction survey received a high score (88%) for an overall satisfaction.
CONCLUSIONS
This device significantly expands the data management and networking capabilities of the bedside glucometer. Networking technology should allow for a centralized quality control management.

Keyword

Glucometer; POCT; DMS (data management system)

MeSH Terms

Clinical Laboratory Information Systems
Glucose
Information Management
Quality Control
Glucose

Cited by  2 articles

Performance Evaluation of BAROZEN H, a Networking Blood Glucose Monitoring System for Medical Institutions
Jun Hyung Lee, Young Joo Cha
Lab Med Online. 2015;5(2):69-76.    doi: 10.3343/lmo.2015.5.2.69.

Analytical Performance Evaluation of Glucose Monitoring System Following ISO15197
Dongheui An, Hee-Jung Chung, Hye-Won Lee, Woochang Lee, Sail Chun, Won-Ki Min
Korean J Lab Med. 2009;29(5):423-429.    doi: 10.3343/kjlm.2009.29.5.423.


Reference

References

1. Handorf CR. Background-setting the stage for alternate-site laboratory testing. Clin Lab Med. 1994; 14:451–8.
Article
2. Worth R, Home PD, Johnston DG, Anderson J, Ashworth L, Burrin JM, et al. Intensive attention improves glycaemic control in insulin dependent diabetes without further advantage from home blood glucose monitoring; results of a controlled trial. British Medical J. 1982; 285:1233–40.
3. National Committee for Clinical Laboratory Standards: Evaluation of precision performance of clinical chemistry devices-2nd ed Teantative guideline. Document EP5-T2.Vilanova, PA: National Committee for Clinical Laboratory Standards;1992.
4. Kilgore ML, Steindel SJ, Smith JA. Evaluating stat testing options in an academic health center: therapeutic turnaround time and staff satisfaction. Clin Chem. 1998; 44:1597–603.
Article
5. Lee SY, Lee SG, Kim JW, Min WK, Park HS. Evaluaion of precision Q.I.DR glucose testing system. Korean J Clin Pathol. 1999; 19:425–32.
6. Lewandrowski E, Mac Millan D, Misiano D, Tochka L, Lewandrowski K. Process improvement for bedside capillary glucose testing in a large academic medical center: the impact of new technology on point-of-care testing. Clin Chim Acta. 2001; 307:175–9.
Article
7. Halpern NA, Brentjens T. Point of care testing informatics. Crit Care Clin. 1999; 15:577–91.
Article
8. Blick KE. The essential role of information management in point-of-care/critical care testing. Clin Chim Acta. 2001; 307:159–68.
Article
9. Cha CH, Lee WC, Lee YW, Kim YC, Kim YH, Chun SI, et al. Installation of data management system for point of care blood gas analyzers using local area network. Korean J Lab Med. 2005; 25:471–6.