Korean J Anesthesiol.  2004 Nov;47(5):764-766. 10.4097/kjae.2004.47.5.764.

Thermal Burn Caused by a Circulating Water Mattress during an Operation: A case report

Affiliations
  • 1Department of Anesthesiology and Pain Medicine, Catholic University of Korea, Seoul, Korea. hongs@catholic.ac.kr

Abstract

A 50-year-old woman underwent left pneumonectomy under general anesthesia. A circulating water mattress covered by a cotton sheet was placed on the operating table, because of an anticipated prolonged surgical time. The mattress's temperature regulator was set at 37oC. After 7 hours of operation in the right lateral decubitus position, 10 x 8 cm sized mixed first- and second-degree burn was found on the right greater trochanteric area, the main weight-bearing site. Although healing occurred without skin graft or complication, this case emphasizes the importance of vigilance to avoid pressure sores and thermal burn injuries during prolonged operations using a circulating water mattress.

Keyword

burn; pressure sore; prolonged operation; circulating water mattress

MeSH Terms

Anesthesia, General
Burns*
Female
Femur
Humans
Middle Aged
Operating Tables
Operative Time
Pneumonectomy
Pressure Ulcer
Skin
Transplants
Water*
Weight-Bearing
Water