Abstract

BACKGROUND: Hypoperfusion during manipulation of the heart in off-pump coronary artery bypass (OPCAB) surgery may lead to postoperative neurological complications. Therefore, it will be necessary to monitor cerebral function during OPCAB surgery. In this study, we compared regional cerebral oxygenation (rSO2) by near-infrared spectroscopy (NIRS) with jugular bulb venous oxygen saturation (SjvO2) and assessed whether rSO2 measured by NIRS could be an alternative method of SjvO2 during OPCAB surgery.
METHODS
A total of 20 patients who underwent OPCAB surgery were studied. A fiberoptic catheter was placed in the right jugular bulb to measure SjvO2 while a cerebral oximeter based on NIRS, INVOS 5100B was used to monitor rSO2. Radial arterial and jugular bulb blood samples were drawn simultaneously from baseline every hour during operation. The values of rSO2 were compared with SjvO2 values.
RESULTS
For all data points (n = 78) for all patients combined, rSO2 values were significantly correlated with SjvO2 values (r = 0.513, P < 0.0001). There were significant correlations between arterial carbon dioxide and values of SjvO2 (r = 0.393, P = 0.0002) and rSO2 (r = 0.432, P < 0.0001).
CONCLUSIONS
We concluded that NIRS correlates with SjvO2 in this patient population. These findings suggest that near-infrared spectroscopy may be useful in assessing cerebral oxygenation during OPCAB surgery.

Keyword

cardiac anesthesia; cerebral oxygenation; jugular bulb venous oxygen saturation; near-infrared spectroscopy; off-pump CABG

MeSH Terms

Carbon Dioxide
Catheters
Coronary Artery Bypass, Off-Pump
Heart
Humans
Organothiophosphorus Compounds
Oxygen
Spectroscopy, Near-Infrared
Carbon Dioxide
Organothiophosphorus Compounds
Oxygen
Full Text Links
  • KJAE
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr