J Korean Ophthalmol Soc.  2014 Nov;55(11):1631-1635. 10.3341/jkos.2014.55.11.1631.

Comparison of Ocular Biometry Measured Using Four Applanation Ultrasonographic Biometry Devices

Affiliations
  • 1Department of Ophthalmology and Visual Science, Bucheon St. Mary's Hospital, The Catholic University of Korea College of Medicine, Bucheon, Korea. eunchol@hanmail.net
  • 2Kangnam St. Mary's LASIK Center, Seoul, Korea.

Abstract

PURPOSE
To compare ocular biometry measured using 4 applanation ultrasonographic biometry devices and evaluate the accuracies of the refractive outcomes after cataract surgery.
METHODS
A total of 60 eyes in 60 patients who received cataract surgery were included in the present study. The axial length was measured using applanation ultrasonographic biometry devices (Aviso(R), Hi-Scan(R), UD-6000(R), P37-II(R)). Additionally, keratometry was measured using an autokeratometer (Topcon KR 8000) and the SRK/T formula was used to calculate intraocular lens (IOL) power. Two months after cataract surgery, the refractive outcome was determined, and results from the 4 different applanation ultrasonographic biometry devices were compared.
RESULTS
Axial lengths were 23.52 +/- 1.45 mm, 23.51 +/- 1.04 mm, 23.54 +/- 1.58 mm, and 23.52 +/- 1.38 mm measured using Aviso(R), Hi-Scan(R), UD-6000(R), and P37-II(R), respectively with no statistically significant differences observed (p = 0.92). The mean absolute error (MAE) of the Aviso(R), Hi-Scan(R), UD-6000(R), and P37-II(R) was 0.41 +/- 0.32 diopter (D), 0.40 +/- 0.30 D, 0.36 +/- 0.26 D, and 0.39 +/- 0.26 D, respectively. The mean numerical error (MNE) was 0.39 +/- 0.37 D, 0.36 +/- 0.32 D, 0.26 +/- 0.29 D, and 0.38 +/- 0.32 D, respectively. The differences between the 4 different applanation ultrasonographic biometry devices were not statistically significant (p = 0.90, p = 0.81).
CONCLUSIONS
The ocular biometric measurements and prediction of postoperative refraction using Aviso(R), Hi-Scan(R), UD-6000(R), P37-II(R) showed no significant differences.

Keyword

Applanation ultrasonographic biometry; A-scan; Axial length; Refractive error

MeSH Terms

Biometry*
Cataract
Humans
Lenses, Intraocular
Refractive Errors

Reference

References

1. Olsen T. Sources of error in intraocular lens power calculation. J Cataract Refract Surg. 1992; 18:125–9.
Article
2. Holzer MP, Mamusa M, Auffarth GU. Accuracy of a new partial coherence interferometry analyser for biometric measurements. Br J Ophthalmol. 2009; 93:807–10.
Article
3. Lam AK, Chan R, Pang PC. The repeatability and accuracy of axial length and anterior chamber depth measurements from the IOLMaster. Ophthalmic Physiol Opt. 2001; 21:477–83.
4. Freeman G, Pesudovs K. The impact of cataract severity on measurement acquisition with the IOLMaster. Acta Ophthalmol Scand. 2005; 83:439–42.
Article
5. Tehrani M, Krummenauer F, Blom E, Dick HB. Evaluation of the practicality of optical biometry and applanation ultrasound in 253 eyes. J Cataract Refract Surg. 2003; 29:741–6.
Article
6. Németh J, Fekete O, Pesztenlehrer N. Optical and ultrasound measurement of axial length and anterior chamber depth for intraocular lens power calculation. J Cataract Refract Surg. 2003; 29:85–8.
Article
7. Olsen T. The accuracy of ultrasonic determination of axial length in pseudophakic eyes. Acta Ophthalmol (Copenh). 1989; 67:141–4.
Article
8. Connors R 3rd, Boseman P 3rd, Olson RJ. Accuracy and reproducibility of biometry using partial coherence interferometry. J Cataract Refract Surg. 2002; 28:235–8.
Article
9. Kielhorn I, Rajan MS, Tesha PM, et al. Clinical assessment of the Zeiss IOLMaster. J Cataract Refract Surg. 2003; 29:518–22.
Article
10. Hennessy MP, Franzco Chan DG. Contact versus immersion biometry of axial length before cataract surgery. J Cataract Refract Surg. 2003; 29:2195–8.
Article
11. Rhim JW, Kang SY, Kim HM. Comparative analysis of contact and immersion technique in ultrasonographic biometry. J Korean Ophthalmol Soc. 2009; 50:1795–9.
Article
12. Binkhorst RD. The accuracy of ultrasonic measurement of the axial length of the eye. Ophthalmic Surg. 1981; 12:363–5.
Article
13. Wirtitsch MG, Findl O, Menapace R, et al. Effect of haptic design on change in axial lens position after cataract surgery. J Cataract Refract Surg. 2004; 30:45–51.
Article
14. Kim SI, Kang SJ, Oh TH, et al. Accuracy of ocular biometry and postoperative refraction in cataract patients with AL-Scan®. J Korean Ophthalmol Soc. 2013; 54:1688–93.
Article
15. Shin JA, Chung SK. Comparison of the refractive results measured by ultrasound and partial coherence interferometers. J Korean Ophthalmol Soc. 2013; 54:723–7.
Article
16. Kwag JY, Choi SH. Comparison of ocular biometry measured by ultrasound and two kinds of partial coherence interferometers. J Korean Ophthalmol Soc. 2011; 52:169–74.
Article
Full Text Links
  • JKOS
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr