J Korean Ophthalmol Soc.  2002 Dec;43(12):2513-2518.

Accuracy of Orbscan Pachymetry Measurements and Ultrasonic Pachymetry before and after LASIK with Orbscan II(R) Topography

Affiliations
  • 1Department of Ophthalmology, Hanil General Hospital, Korea.
  • 2Department of Ophthalmology, Hanyang University Hospital, Korea. Song@email.hanyang.ac.kr

Abstract

PURPOSE: To assess the the accuracy and variability in pachymetry measurements obtained by Orbscan and ultrasonic pachymeter in laser in situ keratomileusis (LASIK).
METHODS
Thirty two eyes of 16 patients before LASIK were measured by ultrasonic pachymeter and Orbscan. Thirty eyes of 16 patients that had LASIK were measured by ultrasonic pachymeter and Orbscan postoperatively. The theoretical residual corneal thickness was compared to measurements by both instruments in eyes that had LASIK procedures.
RESULTS
The ultrasound pachymeter measurements (543.8+/-24.2 micrometer) were thicker than the Orbscan pachymetry measurements, (541.2+/-24.9 micrometer) by mean of 2.7+/-7.4 micrometer (SD), preoperatively. But this is not statistically significant (p>0.05). Postoperatively, the ultrsound measurements were consistent with theoretical residual corneal thickness, 461.8+/-37.4 micrometer versus 467.9+/-30.71 micrometer (p=0.064), while Orbscan measurements were statistically less than the theoretical residual corneal thickness, 433.2+/-48.75 micrometer versus 464.32+/-33.4 micrometer (p<0.001).
CONCLUSIONS
In LASIK eyes, the ultrasonic pachymetry was a more accurate measurement of corneal pachymetry than Orbscan topography system.

Keyword

Corneal thickness; LASIK; Orbscan topography system; Ultrasonic pachymetry

MeSH Terms

Corneal Pachymetry
Humans
Keratomileusis, Laser In Situ*
Ultrasonics*
Ultrasonography