J Korean Ophthalmol Soc.  2002 Jan;43(1):5-10.

Comparison of Corneal Thickness Measurements Between Orbscan Topography System and Ultrasonic Pachymetry before and after LASIK

Affiliations
  • 1Department of Ophthalmology, College of Medicine, University of Ulsan, Asan Medical Center, Korea. hwtchah@www.amc.seoul.kr

Abstract

PURPOSE: To report the difference of the values of the corneal thickness measurement measured by Orbscan topography system and ultrasonic pachymetry is different between pre-LASIK and post-LASIK.
METHODS
The corneal thickness measured by these two methods was compared in 32 pre-LASIK eyes of 18 subjects and 31 post-LASIK eyes of 20 subjects passed 3 months after laser in situ keratomileusis (LASIK).
RESULTS
In the pre-LASIK eyes, the cornea measured thicker by Orbscan topography system than by ultrasonic pachymeter (16.94+/-13.1 microgram, average +/- standard deviation, p=0.001). In the post-LASIK eyes, there was no statistically significant difference in the values of corneal thickness measurement between two methods (-9.58+/-27.88 microgram, p=0.0653).
CONCLUSION
We must give attention to the interpretation of values of corneal thickness measurement measured by Orbscan topography system and ultrasonic pachymetry because the correlations between the two methods in pre and post-LASIK eyes are different. We assume that differences of the values of measurement between the two methods may result from either technical errors of the measurers , interference of corneal interface, or change of tear film.

Keyword

Corneal thickness; LASIK; Orbscan topography system; Ultrasonic pachymetry

MeSH Terms

Cornea
Corneal Pachymetry*
Keratomileusis, Laser In Situ*
Tears
Ultrasonics*
Full Text Links
  • JKOS
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr