Abstract

Fourteen consecutive patients underwent application of V.S.P(Variable Screw Placement) spinal plate between January 1989 and July 1989. Operative indications were spondylolisthesis, spinal stenosis, segmental instability, herniated disc with instability, pseudoarthrosis. unstable fracture, and failed surgery syndrome. Our patients consist of 8 cases of spondylolisthesis, 4 cases of fracture-dislocation, one case of spondylolysis with instability, one case of herniated disc(L1- 2 interspace). Overall results showed 36% excellent, 50% good, 7% fair, 7% poor. Two deep wound infections were observed. Screw alignment and the angular relationship of each screw to the spinal plate is considered as important technical factors in minimizing screw failure.

Keyword

V.S.P. spinal plate; Spondylolisthesis; Unstable spine