J Dent Rehabil Appl Sci.  2015 Dec;31(4):310-315. 10.14368/jdras.2015.31.4.310.

Evaluation of fitness in implant screw as tightening torque in dental laboratory

Affiliations
  • 1Department of Prosthodontics, College of Dentistry, Dankook University, Cheonan, Republic of Korea. ygsong@dankook.ac.kr

Abstract

PURPOSE
The purpose of this study was to measure the tightening torque for dental implant in dental laboratory and to analyze of the effects of different tightening torque.
MATERIALS AND METHODS
The tightening torque for dental implant in dental laboratory were measured by digital torque gauge. The length of abutment and analog were measured as tightening torque of manufacturer's instructions and the measured value. And the data were statically analyzed.
RESULTS
The mean tightening torque of implant screw in dental laboratory was 1.563 +/- 0.332 Ncm. The external type implant system of total length were showing no significant differences but the internal type implant system had difference significant (P < 0.05) when compared with tightening torque.
CONCLUSION
The implant prosthesis should be made under manufacturer's instructions especially as tightening torque of screw. For the fidelity of implant prosthesis, dental technician should learn how to use the torque gauge.

Keyword

implant torque controller; dental technician; implant-abutment; settlement

MeSH Terms

Dental Implants
Dental Prosthesis
Dental Technicians
Humans
Laboratories, Dental*
Prostheses and Implants
Torque*
Dental Implants