J Korean Radiol Soc.  2003 Dec;49(6):507-511. 10.3348/jkrs.2003.49.6.507.

Clinical Image Evaluation of Mammograms: A National Survey

Affiliations
  • 1Department of Radiology, Seoul National University College of Medicine and the Institute of Radiation Medicine, SNUMRC. moonwk@radcom.snu.ac.kr
  • 2Kimmihye Breast Clinic.
  • 3Department of Radiology, Samsung Medical Center, School of Medicine, Sungkyunkwan University.
  • 4Department of Diagnostic Radiology, Radiological Institute of Radiological Science, Yonsei University College of Medicine.
  • 5Department of Diagnostic Radiology, College of Medicine, Ewha Womans University.
  • 6Department of Radiology, Hallym University College of Medicine.
  • 7Department of Diagnostic Radiology, College of Medicine Pochon CHA University.
  • 8Department of Diagnostic Radiology, Gachon Medical School Gil Medical Center.

Abstract

PURPOSE: The goal of this study was to survey the overall quality of mammographic images in Korea.
MATERIALS AND METHODS
A total of 598 mammographic images collected from 257 hospitals nationwide were reviewed in terms of eight image quality categories, namely positioning, compression, contrast, exposure, sharpness, noise, artifacts, and examination identification, and rated on a five-point scale: (1=severe deficiency, 2=major deficiency, 3=minor deficiency, 4=good, 5=best). Failure was defined as the occurrence of more than four major deficiencies or one severe deficiency (score of 1 or 2). The results were compared among hospitals of varying kinds, and common problems in clinical image quality were identified.
RESULTS
Two hundred and seventeen mammographic images (36.3%) failed the evaluation. Poor images were found in descending order of frequency, at The Society for Medical Examination (33/69, 47.8%), non-radiology clinics (42/88, 47.7%), general hospitals (92/216, 42.6%), radiology clinics (39/102, 38.2%), and university hospitals (11/123, 8.9%) (p<0.01, Chi-square test). Among the 598 images, serious problems which occurred were related to positioning in 23.7% of instances (n=142) (p<0.01, Chi-square test), examination identification in 5.7% (n=34), exposure in 5.4% (n=32), contrast in 4.2% (n=25), sharpness in 2.7% (n=16), compression in 2.5% (n=15), artifacts in 2.5% (n=15), and noise in 0.3% (n=2).
CONCLUSION
This study showed that in Korea, 36.3% of the mammograms examined in this sampling had important image-related defects that might have led to serious errors in patient management. The failure rate was significantly higher in non-radiology clinics and at The Society for Medical Examination than at university hospitals.

Keyword

Mammography; Breast radiography, quality assurance

MeSH Terms

Artifacts
Hospitals, General
Hospitals, University
Humans
Korea
Mammography
Noise
Full Text Links
  • JKRS
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr