Clin Orthop Surg.  2015 Mar;7(1):22-28. 10.4055/cios.2015.7.1.22.

The Targon FN System for the Management of Intracapsular Neck of Femur Fractures: Minimum 2-Year Experience and Outcome in an Independent Hospital

Affiliations
  • 1Department of Trauma and Orthopaedics, Queen Elizabeth Hospital, London, UK. donaldosaru@hotmail.com

Abstract

BACKGROUND
The Targon FN implant was developed in 2007 to treat intracapsular neck of femur fractures. Early results from the design centre have shown good results in terms of fracture complications. We wished to see if these results can be reproduced in an independent institution.
METHODS
The records of consecutive patients, treated with this implant between 2008 and 2011 at Queen Elizabeth Hospital, were identified and collected for this study. Operations were performed by all grades of surgeons under supervision as appropriate. These patients went on to have both clinical and radiological assessment for fracture healing and function.
RESULTS
Fifty-one patients were identified with 43 patients available for final follow-up. The average age was 66 years with a minimum follow-up of 24 months. A non-union rate of 0% in the undisplaced fracture group and 1 in 12 (8%) in the displaced fracture group was observed. An avascular necrosis rate of 6% and 8% was observed for undisplaced and displaced fracture types, respectively. No significant change in premorbid to postoperative ambulation was observed and there was no wound complication.
CONCLUSIONS
Our study shows similar results with those of the design centre and which are superior to those currently found in the literature for the more traditional fixation methods. It also shows that the promising results with this new implant as seen from the design institutions can be reproduced by all cadres of surgeons in non-specialist practice.

Keyword

Neck femur; Fracture fixation; Bone nails

MeSH Terms

Adult
Aged
Aged, 80 and over
*Bone Plates
Bone Screws
Female
Femoral Neck Fractures/radiography/*surgery
Follow-Up Studies
Fracture Fixation, Internal/*instrumentation/methods
Humans
Male
Middle Aged
Prosthesis Design
Retrospective Studies
Young Adult

Figure

  • Fig. 1 Radiograph showing Targon FN (B-Braun AG, Melsungen, Germany) implant.

  • Fig. 2 (A) Preoperative radiograph showing undisplaced neck of femur fracture. (B) Radiograph showing fixed undisplaced neck of femur fracture.

  • Fig. 3 (A) Preoperative radiograph showing displaced neck of femur fracture. (B) Radiograph showing fixed displaced neck of femur fracture.

  • Fig. 4 (A) Radiograph showing avascular necrosis of right femoral head. (B) Magnetic resonance imaging confirming avascular necrosis of right femoral head.


Reference

1. National Institute for Health and Clinical Excellence. Hip fracture: the management of hip fractures in adults [Internet]. London: National Institute for Health and Clinical Excellence;2011. cited 2014 Dec 26. Available from: http://www.orthointerview.com/resources/NICE-Hip-Quick-Guide.pdf.
2. Lu-Yao GL, Keller RB, Littenberg B, Wennberg JE. Outcomes after displaced fractures of the femoral neck: a meta-analysis of one hundred and six published reports. J Bone Joint Surg Am. 1994; 76(1):15–25.
3. Parker MJ, Stockton G. Internal fixation implants for intracapsular proximal femoral fractures in adults. Cochrane Database Syst Rev. 2001; (4):CD001467.
4. Parker MJ, Stedtfeld HW. Internal fixation of intracapsular hip fractures with a dynamic locking plate: initial experience and results for 83 patients treated with a new implant. Injury. 2010; 41(4):348–351.
5. Parker MJ. Targon FN®: a new dynamic locking plate and screw system [Internet]. Tuttingen (DE): Aesculap AG;2008. cited 2011 Jun 21. Available from: http://www.aesculap-extra.net/public/frame_doc_index.html?med_id=1000016462.
6. Madsen F, Linde F, Andersen E, Birke H, Hvass I, Poulsen TD. Fixation of displaced femoral neck fractures: a comparison between sliding screw plate and four cancellous bone screws. Acta Orthop Scand. 1987; 58(3):212–216.
7. Linke B, Schwieger K, Bursic D, Dutoit C, Ulrich D, Gautier E. Treatment of unstable femoral neck fractures: is the dynamic hip screw a superior alternative to 3 cannulated screws? Poster session presented at: OTA Annual Meeting (Poster 55). 8-10 October 2004; Hollywood: Orthopaedic Trauma Association;2004.
8. Brandt E, Verdonschot N. Biomechanical analysis of the sliding hip screw, cannulated screws and Targon® FN in intracapsular hip fractures in cadaver femora. Injury. 2011; 42(2):183–187.
9. Conn KS, Parker MJ. Undisplaced intracapsular hip fractures: results of internal fixation in 375 patients. Clin Orthop Relat Res. 2004; (421):249–254.
10. Parker M, Cawley S, Palial V. Internal fixation of intracapsular fractures of the hip using a dynamic locking plate: two-year follow-up of 320 patients. Bone Joint J. 2013; 95B(10):1402–1405.