BACKGROUND/AIMS
It is believed that disposable biopsy forceps are more costly than reusable biopsy forceps. In this study, we evaluated performance and cost of disposable forceps versus reusable forceps in esophagogastroduodenoscopic biopsy.
METHODS
Between October 2009 and July 2010, we enrolled 200 patients undergoing esophagogastroduodenoscopic biopsy at Seoul St. Mary's Hospital. Biopsies were performed with 100 disposable or 5 reusable forceps by random assignment. Seventy-five additional patients were studied to estimate durability of reusable forceps. The assisting nurses estimated the performance of the forceps. The evaluation of costs included purchase prices and reprocessing costs. The adequacy of the sample was estimated according to the diameter of the obtained tissue.
RESULTS
Performance of disposable forceps was estimated as excellent in 97.0%, good in 2.0% and adequate in 1.0%. Reusable forceps were estimated as excellent in 36.0%, good in 36.0%, adequate in 25.1% and inadequate in 2.9%. The performance of reusable forceps declined with the number of uses. The reprocessing cost of reusable forceps for one biopsy session was calculated as 8,021won. The adequacy of the sample was excellent for both forceps.
CONCLUSIONS
Disposable forceps showed excellent performance. Considering the reprocessing costs of reusable forceps, usage of disposable forceps with a low price should be considered.

Keyword

Endoscopy; Biopsy; Disposable equipment; Costs and cost analysis

MeSH Terms

Biopsy
Costs and Cost Analysis
Disposable Equipment
Endoscopy
Humans
Surgical Instruments

Figure

  • Fig. 1 Performance of reusable biopsy forceps according to number of uses.

  • Fig. 2 Mean score of performance of the reusable biopsy forceps according to number of uses.

  • Fig. 3 Cost per use of reusable biopsy forceps with the number of uses.

  • Fig. 4 Size of obtained samples.


Cited by  5 articles

Multidisciplinary and Multisociety Practice Guideline on Reprocessing Flexible Gastrointestinal Endoscopes and Endoscopic Accessories
Dae Young Cheung, Byung Ik Jang, Sang Wook Kim, Jie-Hyun Kim, Hyung Keun Kim, Jeong Eun Shin, Won Jae Yoon, Yong Kang Lee, Kwang Hyun Chung, Soo-Jeong Cho, Hyun Phil Shin, Sun Young Cho, Woon Geon Shin, Kee Don Choi, Byung-Wook Kim, Joong Goo Kwon, Hee Chan Yang, Tae-Geun Gweon, Hyun Gun Kim, Dong-Won Ahn, Kwang Bum Cho, Sun Hee Kim, Kyong Hwa Hwang, Hee Hyuk Im
Clin Endosc. 2020;53(3):276-285.    doi: 10.5946/ce.2020.106.

Korean Society of Gastrointestinal Endoscopy Guidelines for Endoscope Reprocessing
Byoung Kwan Son, Byung-Wook Kim, Won Hee Kim, Dae-Sung Myung, Young-Seok Cho, Byung Ik Jang
Clin Endosc. 2017;50(2):143-147.    doi: 10.5946/ce.2017.029.

Steps of Reprocessing and Equipments
Yong Kook Lee, Jeong Bae Park
Clin Endosc. 2013;46(3):274-279.    doi: 10.5946/ce.2013.46.3.274.

Current Status of Endoscope Reprocessing in Korea
Young-Seok Cho
Clin Endosc. 2015;48(1):1-3.    doi: 10.5946/ce.2015.48.1.1.

Role of Clinical Endoscopy in Emphasizing Endoscope Disinfection
Ji Kon Ryu, Eun Young Kim, Kwang An Kwon, Il Ju Choi, Ki Baik Hahm
Clin Endosc. 2015;48(5):351-355.    doi: 10.5946/ce.2015.48.5.351.


Reference

1. O'Connor HJ, Axon AT. Gastrointestinal endoscopy: infection and disinfection. Gut. 1983; 24:1067–1077. PMID: 6414894.
2. Martin MA, Reichelderfer M. Association for Professionals in Infection Control and Epidemiology, Inc. 1991, 1992, and 1993 APIC Guidelines Committee. APIC guidelines for infection prevention and control in flexible endoscopy. Am J Infect Control. 1994; 22:19–38. PMID: 8172372.
3. Schaffner M. Infection control issues in the gastrointestinal endoscopy unit. Gastroenterol Nurs. 1990; 12:279–284. PMID: 2288941.
Article
4. Cleaning and disinfection of equipment for gastrointestinal flexible endoscopy: interim recommendations of a Working Party of the British Society of Gastroenterology. Gut. 1988; 29:1134–1151. PMID: 3410338.
5. Spach DH, Silverstein FE, Stamm WE. Transmission of infection by gastrointestinal endoscopy and bronchoscopy. Ann Intern Med. 1993; 118:117–128. PMID: 8416308.
Article
6. Bond WW. Virus transmission via fiberoptic endoscope: recommended disinfection. Md Med J. 1988; 37:497–498. PMID: 3172995.
7. Infection control during gastrointestinal endoscopy. Guidelines for clinical application. Gastrointest Endosc. 1988; 34(3 Suppl):37S–40S. PMID: 3384292.
8. Rizzo J, Bernstein D, Gress F. A performance, safety and cost comparison of reusable and disposable endoscopic biopsy forceps: a prospective, randomized trial. Gastrointest Endosc. 2000; 51:257–261. PMID: 10699767.
Article
9. Yang R, Ng S, Nichol M, Laine L. A cost and performance evaluation of disposable and reusable biopsy forceps in GI endoscopy. Gastrointest Endosc. 2000; 51:266–270. PMID: 10699769.
Article
10. Deprez PH, Horsmans Y, Van Hassel M, Hoang P, Piessevaux H, Geubel A. Disposable versus reusable biopsy forceps: a prospective cost evaluation. Gastrointest Endosc. 2000; 51:262–265. PMID: 10699768.
Article
11. Bourguignon C, Destrumelle AS, Koch S, et al. Disposable versus reusable biopsy forceps in GI endoscopy: a cost-minimization analysis. Gastrointest Endosc. 2003; 58:226–229. PMID: 12872090.
12. Bronowicki JP, Venard V, Botté C, et al. Patient-to-patient transmission of hepatitis C virus during colonoscopy. N Engl J Med. 1997; 337:237–240. PMID: 9227929.
Article
13. Tytgat GN. Endoscopic transmission of Helicobacter pylori. Aliment Pharmacol Ther. 1995; 9(Suppl 2):105–110. PMID: 8547522.
14. Mapstone NP, Lynch DA, Lewis FA, et al. Identification of Helicobacter pylori DNA in the mouths and stomachs of patients with gastritis using PCR. J Clin Pathol. 1993; 46:540–543. PMID: 8331177.
Article
15. Bond WW, Moncada RE. Viral hepatitis B infection risk in flexible fiberoptic endoscopy. Gastrointest Endosc. 1978; 24:225–232. PMID: 680530.
Article
16. Yoon JH, Yoon BC, Lee JK, et al. [Comparison of disinfection methods by ethylene oxide gas and steam sterilization of reusable forceps]. Korean J Gastrointest Endosc. 2009; 39(Suppl 2):124S.
Full Text Links
  • CE
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr