Korean J Anesthesiol.  2014 Nov;67(5):342-345. 10.4097/kjae.2014.67.5.342.

Misinsertion of central venous catheter into the suspected vertebral vein: a case report

Affiliations
  • 1Department of Anesthesiology and Pain Medicine, Yeungnam University College of Medicine, Daegu, Korea. apsj0718@naver.com

Abstract

We experienced a case in which a central venous catheter (CVC) was misplaced into the wrong vein, which was mistaken for the internal jugular vein (IJV), identified by chest x-ray and ultrasound. The vertebral vein passes through the transverse foramina from the atlas to the 6th cervical vertebra. After exiting the transverse foramen of the 6th vertebra, the vein subsequently runs anterolateral to the vertebral artery and posterior to the IJV and drains the innominate vein. In this case, chest x-ray and ultrasound revealed that the inserted CVC had a course very similar to the vertebral vein. The misplacement of a CVC into the vertebral vein might occur from excessive rotation of the patient's head, which leads to alterations in the cervical vascular anatomy, and from deep insertion of the puncture needle. Therefore, it is advised, for safe CVC insertion, to minimize a patient's head rotation and to make use of ultrasound when the anatomical structures cannot be clearly identified.

Keyword

Central venous catheters; Jugular veins; Spine

MeSH Terms

Brachiocephalic Veins
Central Venous Catheters*
Head
Jugular Veins
Needles
Punctures
Spine
Thorax
Ultrasonography
Veins*
Vertebral Artery

Cited by  1 articles

Guidewire insertion into the vertebral vein during right internal jugular vein central venous catheterization -A rare case report-
Jeonghan Lee, Jaewoo Suh, Juseok Oh, Seunghee Ki
Anesth Pain Med. 2023;18(4):382-388.    doi: 10.17085/apm.23052.

Full Text Links
  • KJAE
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr