J Korean Neurosurg Soc.  1998 Sep;27(9):1230-1234.

Clinical Comparison of Anterior Cervical Plating System in Acute Cervical Injury

Affiliations
  • 1Department of Neurosurgery, College of Medicine, Yeungnam University, Taegu, Korea.

Abstract

We compare the safety and efficacy of locked and non-locked plating systems with retrospective review of 51 consecutive acute cervical injury patients who underwent anterior cervical plating procedures. The two groups were comparable in demographic details, mean age(locked 40.8, non-locked 37.9), average fusion level(locked 1.53, non-locked 1.67) and fusion rate(locked 100%, non-locked 86.7%). Hardware related complication in locked plate group was seen in one patient(2.78%), but did not required the reoperation. In the non-locked plate group, 4 patients developed screw loosening and, in 1 patient, the inferior screw was found to be 2mm posterior to posterior cortex. Reoperation was needed for this case due to symptomatic myelopathy. In the non-locked plate group instrument related reoperations was done in 5 patients(33.4%) and longer operation time was required than locked system. In this study, anterior cervical fusion with plating was nessessory for stabilizing of acute cervical instability and improving neurological outcome and fusion rate. For this purpose, locked system was theoratically safter and technically easier than non-locked system.

Keyword

Locked plate system; Non-locked plate system; Acute cervical spine injury

MeSH Terms

Humans
Reoperation
Retrospective Studies
Spinal Cord Diseases
Full Text Links
  • JKNS
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr