Ann Clin Microbiol.  2015 Jun;18(2):44-51. 10.5145/ACM.2015.18.2.44.

Performance Evaluation of Anyplex plus MTB/NTM and AdvanSure TB/NTM for the Detection of Mycobacterium tuberculosis and Nontuberculous Mycobacteria

Affiliations
  • 1Department of Laboratory Medicine, Eulji General Hospital, Eulji University College of Medicine, Seoul, Korea. jdchae@eulji.ac.kr

Abstract

BACKGROUND
Polymerase chain reaction (PCR) methods from direct specimen are widely used for the rapid and accurate detection of mycobacteria infection. In this study, we evaluated two domestically developed detection kits for Mycobacterium tuberculosis (MTB) and nontuberculous mycobacteria (NTM) using real-time PCR.
METHODS
A total of 348 samples from patients with suspected tuberculosis were tested with real-time PCR over seven months. We performed real-time PCR using the recently developed Anyplex plus MTB/NTM Detection kit (Seegene) with the CFX 96TM Realtime PCR System (Bio-Rad Laboratories) and the conventional AdvanSure TB/NTM real-time PCR kit (LG Life Sciences) with the SLAN Real-time PCR detection system (LG Life Sciences) to evaluate their performance for detecting MTB and NTM.
RESULTS
The two real-time PCR systems showed 96.8% concordance rate for MTB-positive, NTM-positive, and negative results. Based on culture results, the sensitivity and specificity for the detection of MTB using PCR were 71.0% and 94.9% for Anyplex plus, and 78.1% and 93.9% for AdvanSure, respectively. For the detection of NTM, the sensitivity and specificity were 33.3% and 98.4% for Anyplex plus, and 51.7% and 97.9% for AdvanSure. Both PCR systems showed high MTB positive results in bronchial washing and sputum samples.
CONCLUSION
In detecting MTB and NTM, Anyplex plus MTB/NTM (Seegene) and AdvanSure TB/NTM real-time PCR (LG Life Sciences) showed high concordance rate with each other in all samples. Therefore both detection kits can be used as rapid and reliable detection tool for MTB.

Keyword

Nontuberculous mycobacteria; Real-time PCR; Tuberculosis

MeSH Terms

Humans
Mycobacterium tuberculosis*
Nontuberculous Mycobacteria*
Polymerase Chain Reaction
Real-Time Polymerase Chain Reaction
Sputum
Tuberculosis

Figure

  • Fig. 1. Type distributions of samples according to the tests performed. Abbreviations: BAL, bronchoalveolar lavage; Pleural, pleural fluid; CSF, cerebrospinal fluid, SM, solid media; LM, liquid media.


Cited by  1 articles

Molecular Diagnosis of Tuberculosis
Fariz Nurwidya, Diah Handayani, Erlina Burhan, Faisal Yunus
Chonnam Med J. 2018;54(1):1-9.    doi: 10.4068/cmj.2018.54.1.1.


Reference

1.Joint Committee for the Revision of Korean Guidelines for Tuberculosis Korea Centers for Disease Control and Prevention. Korean Guidelines for Tuberculosis. 2nd ed.Korea Centers for Disease Control and Prevention;2014. p. 30–1.
2.Jang MA., Shin SY., Park S., Seong MW., Park SS., Kim SH. Diagnostic Genetics Subcommittee, The Korean Association of Quality Assurance for Clinical Laboratory. Annual report on the external quality assessment of diagnostic genetics in Korea (2013). J Lab Med Qual Assur. 2014. 36:71–83.
Article
3.Korea Centers for Disease Control and Prevention. Manual of Laboratory Tests for Tuberculosis. Korea Centers for Disease Control and Prevention. 2013. 75–80.
4.Joint Committee for the Revision of Korean Guidelines for Tuberculosis Korea Centers for Disease Control and Prevention. Korean Guidelines for Tuberculosis. 2nd ed.Korea Centers for Disease Control and Prevention;2014. p. 227–8.
5.Shinnick TM., Iademarco MF., Ridderhof JC. National plan for reliable tuberculosis laboratory services using a systems approach. Recommendations from CDC and the Association of Public Health Laboratories Task Force on Tuberculosis Laboratory Services.
6.Centers for Disease Control and Prevention (CDC). Updated guidelines for the use of nucleic acid amplification tests in the diagnosis of tuberculosis. MMWR Morb Mortal Wkly Rep. 2009. 58:7–10.
7.Broccolo F., Scarpellini P., Locatelli G., Zingale A., Brambilla AM., Cichero P, et al. Rapid diagnosis of mycobacterial infections and quantitation of Mycobacterium tuberculosis load by two real-time calibrated PCR assays. J Clin Microbiol. 2003. 41:4565–72.
8.Chang HE., Heo SR., Yoo KC., Song SH., Kim SH., Kim HB, et al. Detection of Mycobacterium tuberculosis complex using real-time polymerase chain reaction. Korean J Lab Med. 2008. 28:103–8.
9.Lim J., Kim J., Kim JW., Ihm C., Sohn YH., Cho HJ, et al. Multicenter evaluation of Seegene Anyplex TB PCR for the detection of Mycobacterium tuberculosis in respiratory specimens. J Microbiol Biotechnol. 2014. 24:1004–7.
10.Lee JH., Kim BH., Lee MK. Performance evaluation of anyplex plus MTB/NTM and MDR-TB detection kit for detection of mycobacteria and for antituberculosis drug susceptibility test. Ann Clin Microbiol. 2014. 17:115–22.
Article
11.Lee H., Park KG., Lee G., Park J., Park YG., Park YJ. Assessment of the quantitative ability of AdvanSure TB/NTM real-time PCR in respiratory specimens by comparison with phenotypic methods. Ann Lab Med. 2014. 34:51–5.
Article
12.Cho SY., Kim MJ., Suh JT., Lee HJ. Comparison of diagnostic performance of three real-time PCR kits for detecting Mycobacterium species. Yonsei Med J. 2011. 52:301–6.
Full Text Links
  • ACM
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr