J Korean Acad Prosthodont.
1999 Dec;37(6):825-834.
A COMPARATIVE STUDY ON THE FIT IN PROSTHESES USING PREMADE GOLD CYLINDER AND PLASTIC CYLINDER
Abstract
- Recently, various implant cylinders were supplied, especially received gold cylinders, cast cylinders produced from premade gold and plastic cylinders and plastic cylinders. This study measured and compared precise fit produced when using conventional gold and plastic cylinders. The comparative fit of lapped and non-lapped castings made from plastic pattern was examined. The implant/abutment interface fit was evaluated by scanning electron microscopy(SEM) for each of four cylinders. The following results were obtained: 1. The case of plastic cylinder showed 9.67+/-1.50micrometer gap Steri-Oss fixture was connected. 2. The case oflapped casting made from plastic cylinder showed 3.01+/-2.81micrometer gap when Steri-Oss fixture was connected. 3. The case of gold/plastic cylinder showed 9.80+/-1.68micrometer gap when Steri-Oss fixture was connected. 4. The case of gold cylinder showed 5.47+/-2.43micrometer gap when Steri-Oss fixture was connected. 5. In case of each cylinder which was connected with Steri-Oss fixture, the size of gap was showed less in the order of the case of gold/plastic cylinder, the case of plastic cylinder, the case of gold cylinder and the case of lapped casting made from plastic cylinder. As results of this study, the use of premade gold cylinder offers an advantage over plastic patterns in precise fit. When plastic patterns are used, polishing of implant cylinder components should provide precise fit.