Korean J Anesthesiol.  1987 Oct;20(5):623-626. 10.4097/kjae.1987.20.5.623.

Clinical Application of the Mapleson B System for Controlled Ventilation in Pediatric Patients

Affiliations
  • 1Department of Anesthesiology, National Medical Center, Seoul, Korea.

Abstract

This study was undertaken to observe whether an adult ventilator with a preset volume could be used as a controlled ventrolled ventilator far pediatric anesthesia. 35 Patients ranging in age from 3months to 7 years were divided into two groups based on body weight(Group 1: 5~10kg, 14 cases, Group 2: 11~15kg, 21 cases) and anesthetized with halothane-N2O/O2 - pancuronium using the Mapleson B system. Immediately after induction, the reservoir bay of the Mapleson B system was replaced by the reservoir tube of the adult ventilator (MCM 801). Arterial blood gas studies 30 and 60 minutes after induction were performed, and the data from group 1 was compared with that of group 2. The magnitude of PCO2 increase 30 minutes after induction was not significantly different from that at 60 minutes(p>0.05), and alterations of PCO2 in group 1 were not stati-stically significant with group 2 (P>0.05). In conclusion, it is suggested that the Mapleson B system attached to adult ventilator is an useful and convenient device for controlled ventilation in pediatric patients.


MeSH Terms

Adult
Anesthesia
Bays
Humans
Pancuronium
Ventilation*
Ventilators, Mechanical
Pancuronium
Full Text Links
  • KJAE
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr