Korean J Urol.  2007 Jan;48(1):77-81. 10.4111/kju.2007.48.1.77.

Discrepancy of Measured Renal Length between Ultrasonography and Dimercaptosuccinic Acid (DMSA) Scintigraphy

Affiliations
  • 1Department of Urology, Urological Science Institute, Yonsei University College of Medicine, Seoul, Korea. swhan@yumc.yonsei.ac.kr

Abstract

PURPOSE: An inter-observer variability exists in the measurement of the renal length during ultrasonography, but dimercaptosuccinic acid (DMSA) scintigraphy allows greater objectivity. Thus, the aim of this study was to evaluate the discrepancy of the renal length measured using the two methods.
MATERIALS AND METHODS
Between January 2003 and June 2005, 249 consecutive patients underwent DMSA scintigraphy and ultrasonography. Due to the exclusion of kidneys where a photon defect was observed during DMSA scintigraphy, the number of kidney enrolled in this study was 439. The renal length measurements in any plane were longest using ultrasonography. The renal lengths were measured using DMSA scintigraphy by assessing the maximum pixel length, which was then converted to centimeters.
RESULTS
The mean renal lengths measured by DMSA scitigraphy and ultrasonography were 8.06+/-1.49 and 6.80+/-1.38cm, respectively. The mean difference in the renal length measured between the two methods was 1.26+/-0.84cm. The sonographic renal length was significantly shorter than that found via scintigraphy (Student's t-test, p<0.001).
CONCLUSION
Our results were similar to those of previously reported studies, where the scintigraphic renal length exceeded that obtained sonographically. Due to the possibility of underestimating the renal length when using ultrasonography, clinicians must be cautious when following up pediatric patients using ultrasonography.

Keyword

Kidney; Length; Ultrasonography; Technetium-99m-dimercaptosuccinic acid

MeSH Terms

Humans
Kidney
Observer Variation
Radionuclide Imaging*
Succimer*
Technetium Tc 99m Dimercaptosuccinic Acid
Ultrasonography*
Succimer
Technetium Tc 99m Dimercaptosuccinic Acid

Figure

  • Fig. 1. Measurement of the renal length. (A) Ultrasonography and (B) dimercaptosuccinic acid (DMSA) scintigraphy. Three one centimeter whitish circles are calibration factors. (A) The measured renal lengths in any plane were longest using ultrasonography. (B) Renal lengths were measured using DMSA scintigraphy by assessing the maximum pixel length, which was then converted to centimeters. Scintigraphic

  • Fig. 2. The distribution of the measured renal lengths between DMSA scintigraphy and ultrasonography (r2=0.694, p< 0.001).


Reference

1.Schlesinger AE., Hernandez RJ., Zerin JM., Marks TI., Kelsch RC. Interobserver and intraobserver variations in sonographic renal length measurements in children. AJR Am J Roentgenol. 1991. 156:1029–32.
Article
2.Sargent MA., Wilson BP. Observer variability in the sonographic measurement of renal length in childhood. Clin Radiol. 1992. 46:344–7.
Article
3.Ablett MJ., Coulthard A., Lee RE., Richardson DL., Bellas T., Owen JP, et al. How reliable are ultrasound measurements of renal length in adults? Br J Radiol. 1995. 68:1087–9.
Article
4.Emamian SA., Nielsen MB., Pedersen JF., Ytte L. Kidney dimensions at sonography: correlation with age, sex and habitus in 665 adult volunteers. AJR Am J Roentgenol. 1993. 160:83–6.
Article
5.Hodson CJ., Drewe JA., Kam MN., King A. Renal size in normal children: a radiogrphic study during life. Arch Dis Child. 1962. 37:616–22.
6.Emamian SA., Nielsen MB., Pedersen JF. Intraobserver and interobserver variations in sonographic measurements of kidney size in adult volunteers. A comparison of linear measurement and volumetric estimates. Acta Radiol. 1995. 36:399–401.
7.Carrico CW., Zerin JM. Sonographic measurement of renal length in children: does the position of the patient matter? Pediatr Radiol. 1996. 26:553–5.
Article
8.De Sanctis JT., Connolly SA., Bramson RT. Effect of patient position on sonographically measured renal length in neonates, infants and children. AJR Am J Roentgenol. 1998. 170:1381–3.
Article
9.Fembach SK., Davis TM. The abnormal renal axis in children with spina bifida and gibbus deformity-the pseudohorseshoe kidney. J Urol. 1986. 136:1258–60.
10.Gross GW., Boal DK. Sonographic assessment of normal renal size in children with myelodysplasia. J Urol. 1988. 140:784–6.
Article
11.Farrant P., Meire HB. Ultrasonic measurement of renal inclination; its importance in measurement of renal length. Br J Radiol. 1978. 51:628–30.
Article
12.Rosenbaum DM., Komgold E., Teele RL. Sonographic assessment of renal length in normal children. AJR Am J Roentgenol. 1984. 142:467–9.
Article
13.Wallin L., Thome J., Palmer J., Bajc M. Kidney size estimation in piglets using dimercaptosuccinic acid (DMSA) scintigraphy. Clin Physiol. 1997. 17:591–7.
14.Sisayan RM., Rossleigh MA., Mackey DW. Nomograms of renal length in children obtained from DMSA scintigraphy. Clin Nucl Med. 1993. 18:970–3.
Article
15.Rossleigh MA., Farnsworth RH., Leighton DM., Yong JL., Rose M., Christian CL. Technetium-99m dimercaptosuccinic acid scintigraphy studies of renal cortical scarring and renal length. J Nucl Med. 1998. 39:1280–5.
Full Text Links
  • KJU
Actions
Cited
CITED
export Copy
Close