Korean J Anesthesiol.  2011 Dec;61(6):511-514. 10.4097/kjae.2011.61.6.511.

Inappropriate elevation of bispectral index values in robot assisted thyroidectomy with electromyographic endotracheal tube: A case report

Affiliations
  • 1Department of Anesthesiology and Pain Medicine, College of Medicine, Hallym University, Anyang, Korea. drberry@hanmail.net

Abstract

We report a case of an erroneously elevated bispectral index (BIS) during robot assisted thyroidectomy using an electromyographic endotracheal tube (EMG tube), which is safe and useful for laryngeal electromyographic monitoring. Ten minutes after start of the operation, a sudden increase of BIS value up to 98 was noticed. The BIS values were not decreased to < 65 with supplemental anesthetics. The anesthetic method was changed from total intravenous anesthesia to balanced anesthesia. The BIS sensor and monitor were changed and other models were used. These interventions did not alter BIS values. BIS levels remained between 60 and 70 throughout the main procedure and intermittently increased to the mid-90s without any trace of poor signal quality. At the end of the surgery, the BIS values returned to normal range. The patient did not complain of intraoperative recall. Knowledge of potential interference from the use of an EMG tube must be considered when interpreting BIS.

Keyword

Bispectral index; Electromyographic endotracheal tube; Laryngeal nerve monitoring; Thyroid surgery

MeSH Terms

Anesthesia, Intravenous
Anesthetics
Balanced Anesthesia
Humans
Organothiophosphorus Compounds
Reference Values
Thyroidectomy
Anesthetics
Organothiophosphorus Compounds

Cited by  3 articles

Comparative Analysis of Phase Lag Entropy and Bispectral Index as Anesthetic Depth Indicators in Patients Undergoing Thyroid Surgery with Nerve Integrity Monitoring
Kwon Hui Seo, Kyung Mi Kim, Soo Kyung Lee, Hyunji John, Junsuck Lee
J Korean Med Sci. 2019;34(20):.    doi: 10.3346/jkms.2019.34.e151.

Bispectral index monitoring correlates with the level of consciousness in brain injured patients
Jin Yong Jung, Cheol Beom Cho, Bo Mi Min
Korean J Anesthesiol. 2013;64(3):246-250.    doi: 10.4097/kjae.2013.64.3.246.

Influence of electrocautery-induced electromagnetic interference on quantitative electroencephalographic monitoring of hypnosis during general anesthesia: comparison between the ADMS® and the BIS VISTATM
Dae-hee Kim, Ji-young Yoo, Jong-Yeop Kim, Soo-hwan Ahn, Seongsu Kim, Sang-kee Min
Korean J Anesthesiol. 2018;71(5):368-373.    doi: 10.4097/kja.d.18.27154.

Full Text Links
  • KJAE
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr