BACKGROUND
Chlamydia trachomatis is common etiological agent of sexual transmitted disease. We evaluated two real-time PCR assays for the detection of C. trachomatis; Cobas TaqMan CT test v2.0 (Roche Diagnostics, Switzerland) and AdvanSure CT/NG real-time PCR (LG Life Sciences, Korea). And we compared performance of two real-time PCR assays with Cobas Amplicor CT test (Roche Diagnostics, Switzerland).
METHODS
Seventy-nine cervical swab specimens collected from patients with suspected C. trachomatis infection were tested using CTM CT, ADVANSURE, and AMPLICOR for the detection of C. trachomatis.
RESULTS
Twenty-five (31.6%) samples showed positive results in all three assays. The positive and negative concordance rates among three assays were 100%.
CONCLUSIONS
CTM CT and ADVANSURE showed excellent agreement with AMPLICOR. Therefore, these assays were reliable for the evaluation of C. trachomatis infection.

Keyword

Chlamydia trachomatis; Real-time polymerase chain reaction; Comparison

MeSH Terms

Biological Science Disciplines
Chlamydia
Chlamydia trachomatis
Humans
Real-Time Polymerase Chain Reaction
Full Text Links
  • JLMQA
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr