J Korean Orthop Res Soc.  2004 Oct;7(2):145-151.

Evaluation of Porosity in Cylindrical Bone Cement Specimen Using Image Analysis System

Affiliations
  • 1Department of Orthopaedics, St. Vincent's Hospital, The Catholic University of Korea. Korea. hskoh@unitel.co.kr
  • 2Department of Orthopaedics, St. Mary's Hospital, The Catholic University of Korea.

Abstract

PURPOSE
The porosity of the bone cement is the most important cause of fatigue failure, the most common mode of failure of bone cement using widely in arthroplasty. It is important to evaluate the porosity of bone cement for improvement or development of bone cement, but the conventional 'stain, 'cut, and 'polish, manual method takes long time and efforts. So it is necessary to develop a new technique for evaluation of porosity of bone cement. We tried a technique using computer image analysis system to evaluate the porosity of bone cement specimen and assess efficacy of the method. Simultaneously we evaluated the relationship between the porosity of bone cement and fatigue failure. MATERIAL AND METHODS: We made 59, 2.5inch-length bone cement specimens(30 Simplex P, 29 Palacos R) using Simplex P and Palaces R which are widely using in clinical situation and checked radiogram using mammography film. After scanning the mammography film, we measured the porosity of the bone cement specimens using NIH(National Institute for Health) Image 1.6 version image analysis program. We also, measured the porosity of the bone cement specimens with conventional 'stain','cut' and 'polish' method, after then compared the results of two methods. Simultaneously, we evaluated the relationship between porosity & fatigue failure by loading 9.0, 10.0, 12.5 and 15.0 MPa load with frequency of 10Hz to the bone cement specimens under the physiologic condition.
RESULTS
The coefficient of relation of simplex P and palaces R was 0.729 and 0.713 respectively, so there was high relationship between the image analysis system method and conventional one. It was easy and took shorter time to measure the porosity of bone cement specimens with image analysis system. There was high correlation between cement porosity and fatigue failure, regardless of level of load.
CONCLUSION
It was very easy and fast to measure the porosity of the bone cement specimens with image analysis system and there was high correlation between cement porosity and fatigue failure.

Keyword

Cement Porosity; Image Analysis System; Fatigue Failure

MeSH Terms

Arthroplasty
Fatigue
Mammography
Methylmethacrylate
Porosity*
Methylmethacrylate
Full Text Links
  • JKORS
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr